Performance Test Option Wcdma Demodulation Hardware - R&S Fsp-B15; Checking Comb Line Level - Rohde & Schwarz FSP3 Service Manual

Spectrum analyzer
Table of Contents

Advertisement

R&S FSP
Performance Test Option WCDMA Demodulation Hardware –
R&S FSP-B15

Checking Comb Line Level

R&S FSP settings:
Measurement:
1164.4556.82
- [ PRESET ]
- [ SETUP : SERVICE : NEXT : CAL GEN COMB 62.5 kHz ]
- [ FREQ : CENTER : 40.0625 MHz ]
- [ SPAN : 100 kHz ]
- [ AMPT : - 40 dBm ]
- [ AMPT : RF ATTEN MANUAL : 10 dB ]
- [ TRACE : AVERAGE ]
- [ SWEEP : SWEEP COUNT : 20 ENTER ]
Set marker to peak:
- [ MKR
: PEAK ]
- [ FREQ : CENTER : 50.0625 MHz ]
- [ MKR
: PEAK ]
The signal level L
is displayed by the level reading of marker 1.
FSP
1.27
Performance Test
E-1