Checking the Rated Specifications
5.1.1
Test Setup for Checking the Option NRT-B1
Test setup:
Meßsender
Leistungs-
verstärker
(SMGL)
5.2
Test Procedure
After power up the NRT checks all sensor interfaces for sensors being connected and all options. If a
fault is detected, an error message or warning is displayed. Faults are located according to Section 4.4,
Trouble Shooting. Further tests can be triggered in the UTILity - TEST menu (Section 2.5.11, Test (Built-
in Tests)).
5.2.1
Switch-On Sequence
½ Connect the NRT without sensors to the AC supply.
Preparation:
Test:
The NRT should be automatically switched on without actuation of the ON/STBY
key.
½ Switch off and restart the NRT using the ON/STBY key.
5.2.2
Display and Keypad
5.2.2.1
Display Test
½ Select the function parameter LCD in the UTIL - TEST menu.
Preparation:
Test:
All segments in the display should be shown until the SEL(LOCAL) key is pressed
again.