Rohde & Schwarz FSP3 Service Manual page 46

Spectrum analyzer
Table of Contents

Advertisement

Performance Test
Measurement
Step attenuator settings:
R&S FSP settings:
Evaluation:
10 dB gain steps:
a
10 dB
ATT
Reference level
0 dBm
1 dB gain steps:
a
20 dB
ATT
Reference level
-10 dBm
1164.4556.82
Attenuation
Refer to table below for values of a
- [ AMPT : {reference level} dBm ]
Refer to table below for values of reference level.
- [ MKR
: PEAK ]
The difference between the level of the input signal of the R&S FSP
and the reference (at 10 dB IF gain) is displayed in the marker field by
the reading ´Delta [T1 FXD]´.
20 dB
30 dB
40 dB
-10 dBm
-20 dBm
-30 dBm
21 dB
22 dB
23 dB
-11 dBm
-12 dBm
-13 dBm
{a
}
ATT
.
ATT
50 dB
60 dB
-40 dBm
-50 dBm
24 dB
25 dB
26 dB
-14 dBm
-15 dBm
-16 dBm
1.18
R&S FSP
27 dB
28 dB
29 dB
-17 dBm
-18 dBm
-19 dBm