Completing the Instrument and Frequency Response Correction
Put the 13/30 GHz MW converter unit (135/140) back into the instrument.
Note:
See description "Replacing MW Converter Unit A160".
Execute frequency response correction software R&S FSP-FRQ.EXE.
The software is located on the Gloris server.
Test setup according to menu item "Schematic" (connect external reference).
-
Check IEC/IEEE addresses and configuration of instruments.
-
Press the Autoselect button and perform the measurement with Run.
-
The frequency correction of the YIG filter is performed under item YIG.
-
After calibration, the frequency response is corrected over the entire frequency range.
-
Create a backup of the EEPROM data (see chapter 2, section "Adjustment of Module Data").