Rohde & Schwarz FSP3 Service Manual page 30

Spectrum analyzer
Table of Contents

Advertisement

Measurement Equipment
Item
Type of
equipment
5
Signal generator
6
Power splitter
2-resistor design
7
50
termination
8
Power meter
9
Power sensor
10
Power sensor
11
Step attenuator
12
Fixed attenuator
(2 x)
14
N cable
15
Spectrum
analyzer
16
Arbitrary
waveform
generator
1164.4556.82
Specifications recommended
R&S FSP 07: 10 MHz to 7 GHz
R&S FSP 13: 10 MHz to 13.6 GHz
R&S FSP 30: 10 MHz to 30 GHz
R&S FSP 40:
1 MHz to 40 GHz
R&S FSP 03:
1 MHz to 3 GHz
R&S FSP 07: 10 MHz to 7 GHz
R&S FSP 13: 10 MHz to 13.6 GHz
R&S FSP 30: 10 MHz to 30 GHz
R&S FSP 40: 10 MHz to 40 GHz
Level imbalance
1 MHz to 3 GHz
0.15 dB
3 GHz to 7 GHz
0.2 dB
7 GHz to 13.6 GHz
0.25 dB
13.6 GHz to 40 GHz
0.4 dB
R&S FSP 03:
to 3 GHz
R&S FSP 07:
to 7 GHz
R&S FSP 13:
to 13.6 GHz
R&S FSP 30:
to 30 GHz
R&S FSP 40:
to 40 GHz
1 MHz to 3 GHz
RSS
0.8%
Meter noise
20 pW
RSS referenced to indic. power:
1 MHz to 1 GHz
1.5 %
1 GHz to 7 GHz
2 %
7 GHz to 13.6 GHz
3.5 %
13.6 GHz to 30 GHz
4 %
R&S FSP 03: 10 MHz to 3 GHz
R&S FSP 07: 10 MHz to 7 GHz
R&S FSP 13: 10 MHz to 13.6 GHz
R&S FSP 30: 10 MHz to 30 GHz
R&S FSP 40: 10 MHz to 40 GHz
Variable attenuation
0 dB to 100 dB, 1 dB steps
Attenuation accuracy
< 0.1 dB (f = 128 MHz)
Fixed attenuation 10 dB
R&S FSP 03: 10 MHz to 3 GHz
R&S FSP 07: 10 MHz to 7 GHz
R&S FSP 13: 10 MHz to 13.6 GHz
R&S FSP 30: 10 MHz to 30 GHz
R&S FSP 40: 10 MHz to 40 GHz
Attenuation < 0.2 dB to 3 GHz
Frequency range to 3 GHz
Frequency range to 10 MHz
2 sinusoidal signals with 90 deg.
phase difference
Equipment
Alternative
recommended
equipment
R&S SMP02
HP 63620 B
R&S SMP02
HP 63620 B
R&S SMP04
HP 83640 B
R&S SMP04
HP 83640 B
(Agilent)
HP 11667A
(Agilent)
HP 11667C
+ adapter to K
(Agilent)
RNA
85138A
RNA
(Agilent)
Wiltron 28S50
Wiltron 28K50
Wiltron 28K50
R&S NRVD
HP 483A
(Agilent)
R&S NRV-Z4
HP 8482 A
(Agilent)
HP 8487 A
R&S NRV-Z4
(Agilent)
R&S NRV-Z2
R&S NRV-Z2
R&S NRV-Z55
R&S NRV-Z55
R&S RSP
11713A with
8494G and
8496G
(Agilent)
8490D-010
DNF
(Agilent)
DNF
Wiltron
43KB-10
Wiltron
43KC-10
Wiltron
43KC-10
R&S FSP 3
E4404B
ESA-E
(Agilent)
R&S ADS
33220A with
option 001
(Agilent)
(2 units)
1.2
R&S FSP
Use
Third order intercept
Third order intercept
Frequency response
Noise display
Frequency response
Frequency response
Frequency response
Reference level switching
display linearity
RF attenuator
Third order intercept
TG output level
TG modulation
TG modulation
E-1

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents