Rohde & Schwarz FSP3 Service Manual page 192

Spectrum analyzer
Table of Contents

Advertisement

Troubleshooting
Signal Paths on the IF Filter Board
The RF attenuator is used to provide an appropriate test level. Therefore, a fault in the RF attenuator will
entail subsequent errors with the IF filter test.
Normal action
Input Level of IF Filter / Calibration Amplifier
(CAL-Amps 1 + 2)
Selftest detector A (see block diagram)
The level applied to the RF input is referred to as
mixer level in the following. The RF attenuation is
to be set to 0 dB. The input frequency is always
128 MHz.
The input level at X131 is 8 dB ±3 dB higher than
the mixer level. With maximum mixer level of
-10 dBm the level at X131 is -2 dBm ± 3 dB.
CAL_Amp2 is provided for frequency response
correction. With the input frequency of 128 MHz,
the gain of CAL_Amp2 is set to 0 dB. The
frequency for the selftest is 128 MHz.
The nominal gain between X131 and X132 is
+2 dB.
The level at X132 is 0 dBm with a mixer level of
-10 dBm.
The selftest (A) checks the level to ±5 dB (User)
or ±3 dB (Service Level 1) after passing the
CAL_Amps.
1164.4556.82
Error and error cause
FATAL ERROR!
IF Board: IF input level / CALAMP
Selftest aborted
Possible error causes:
Signal path interrupted in the frontend.
Erroneous EEPROM data in the frontend
leading to incorrect setting of CAL_Amp1.
CAL_Amp1 or 2 faulty.
Troubleshooting
Check the level applied at X131 with
mixer level -10 dBm: -2 dBm, production
tolerance ± 3 dB, maximum tolerance
permitted ± 4.5 dB;
Replace the frontend, if the deviation exceeds
this value
3.88
R&S FSP
E-1

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents