Rohde & Schwarz FSP3 Service Manual page 149

Spectrum analyzer
Table of Contents

Advertisement

R&S FSP
Completing the Instrument and Frequency Response Correction
Put the 13/30 GHz MW converter unit (135/140) back into the instrument.
Note:
See description "Replacing MW Converter Unit A160".
Execute frequency response correction software R&S FSP-FRQ.EXE.
The software is located on the Gloris server.
Test setup according to menu item "Schematic" (connect external reference).
-
Check IEC addresses and configuration of instruments.
-
Press the Autoselect button and perform the measurement with Run.
-
The frequency correction of the YIG filter is performed under item YIG.
-
After calibration, the frequency response is corrected over the entire frequency range.
-
1164.4556.82
3.45
Module Replacement
E-1

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents