General Operating Information; To Set Screen Intensity; To Set Rf Voltage Interpretation (50 /Emf); To Set The Date And Time - HP 8920A User Manual

Rf communications test set
Hide thumbs Also See for 8920A:
Table of Contents

Advertisement

Chapter 2, Configuring Your Test Set

General Operating Information

General Operating Information

To Set Screen Intensity

To Set RF Voltage Interpretation (50 /emf)

To Set the Date and Time

50
S:\HP892XX\USRGUIDE\manual\configts.fb
The following configuration information discusses general operating information
for some of the fields in these screens.
1. Access the CONFIGURE screen.
2. Select the Intensity field.
3. Rotate the knob to change the setting (1=dim, 8=bright).
1. Access the CONFIGURE screen.
2. Position the cursor in front of the RFGen Volts field.
3. Press the Cursor Control knob or press ENTER to select 50 ohm or emf.
Voltage settings can control either:
the voltage across a 50-ohm load, or
the open circuit voltage (emf).
This setting affects the RF Generator's and the Tracking Generator's amplitudes.
1. Access the CONFIGURE screen.
2. Select the Date field and use the DATA keys to enter the date in the format shown be-
low the field.
3. Select the Time field and use the DATA keys to enter the time in the format shown
below the field.
The Test Set has a built-in clock that keeps track of the date and time. It is
powered by an internal battery to keep it operating when the instrument is off.

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents