Chapter 6, Call Processing Subsystem
The CALL CONTROL Screen
TX Power
138
S:\HP892XX\USRGUIDE\manual\callscrn.fb
The
field is displayed only when the
TX Power
This field displays the measured RF power of the RF carrier being transmitted by
the mobile station. A non-zero value would only be displayed when the Test Set's
annunciator is lit).
Connect
When the CALL CONTROL screen is displayed and the Test Set is in the
connected state (
Connect
carrier power is constantly monitored. If the power falls below 0.0005 Watts, the
error message
RF Power Loss indicates loss of Voice Channel
be displayed, and the Test Set will terminate the call and return to the active state
(
annunciator is lit).
Active
In order to ensure that the Test Set makes the correct decisions regarding the presence of
the mobile stations's RF carrier, the Test Set's RF power meter should be zeroed before
using the Call Processing Subsystem. Failure to zero the power meter can result in errone-
ous RF power measurements. See
page 141
for information on manually zeroing the RF Power meter.
Display
annunciator is lit), the mobile station's transmitted
"Conditioning the Test Set for Call Processing" on
field is set to
.
Meas
will