Encoder:mpt1327:Filler:update; Encoder:mpt1327:Filler:clear N; Forward Control Channel In Control Mode And Not Sending - HP 8920A User Manual

Rf communications test set
Hide thumbs Also See for 8920A:
Table of Contents

Advertisement

Figure 102

Forward Control Channel in Control Mode and Not Sending

If the signaling command has the data codeword extension,
subsequent timeslot locations are overwritten with the data codewords. If this
results in a write to a timeslot greater than 32, the write will wrap back to timeslot
one.
A new filler pattern typically comprises several commands to be written. Since
each is written individually, the following update command is also needed to
transfer data into the working filler buffer:

ENCoder:MPT1327:FILLer:UPDATe

To complete the command set for managing the filler buffer, the individual
timeslot locations of the filler buffer are cleared by the following command:

ENCoder:MPT1327:FILLer:CLEAR n

Where n selects the timeslot to be cleared and has the range 1 through 32.
The whole buffer is cleared by the command:
ENCoder:MPT1327:FILLer:RESET
When the
Test Mode
hardware will cycle through the defined part of the filler buffer, transmitting each
128-bit timeslot in turn. This is shown in
shows the control message buffer for comparison with
TIMESLOT 1
2
3
4
7492AB40 . . 00
5
6
7
8
31
32
FILLER BUFFER
Chapter 10, Signaling Encoder (AF Generator 2)
field is in
mode, the signaling firmware/
Control
figure 102 on page 411
7721B4C0 . . D9
5439AB12 . . 01
098BB27A . . 01
201759BC . . A8
91032000 . . 1B
MPT 1327 Encoder
or
DCW2
DCW4
.
Figure 102
figure 103 on page 413
666430CB . . AD
449DDE30 . . 02
67230CB2 . . 04
0980700D . . FF
CC MESSAGE BUFFER
,
also
.
411