Download Print this page

Tools Required For The Installation - Keysight N5231A Installation Note

Configurable test set and source attenuators upgrade kit option 200 to option 216

Advertisement

Getting Prepared
Protecting Your Workspace from Electrostatic Discharge
For information, click on the Chapter 1 bookmark, "Electrostatic Discharge
Protection" in the PDF Service Guide
ESD Equipment Required for the Installation
Description
ESD grounding wrist strap
5-ft grounding cord for wrist strap
2 x 4 ft conductive table mat and 15-ft grounding wire
ESD heel strap (for use with conductive floors)

Tools Required for the Installation

Description
T-10 TORX driver - set to 9 in-lbs (1.02 N.m)
T-20 TORX driver - set to 21 in-lbs (2.38 N.m)
5/16-in (8 mm) nutsetter or open end torque wrench - set to
10 in-lbs (1.13 N.m)
Use a 5/16-in torque wrench set to 10 in-lbs on all cable connections
except the front panel and rear panel bulkhead connectors. Torque these
connections to 21 in-lb.
About Installing the Upgrade
Products affected
Installation to be performed by
Estimated installation time
Estimated adjustment time
Estimated full instrument
calibration time
8
1
.
N5231A/B, N5232A/B, and N5239A/B Option 200
Keysight service center or personnel qualified by
Keysight
2 hours
0.5 hours
4.5 hours
Keysight Part Number
9300-1367
9300-0980
9300-0797
9300-1308
Qty
Part Number
1
N/A
1
N/A
1
N/A
Installation Note N5232-90101

Advertisement

loading

This manual is also suitable for:

N5231bN5232aN5232bN5239aN5239b