Getting Prepared
Protecting Your Workspace from Electrostatic Discharge
For information, click on the Chapter 1 bookmark, "Electrostatic Discharge
Protection" in the PDF Service Guide
ESD Equipment Required for the Installation
Description
ESD grounding wrist strap
5-ft grounding cord for wrist strap
2 x 4 ft conductive table mat and 15-ft grounding wire
ESD heel strap (for use with conductive floors)
Tools Required for the Installation
Description
T-10 TORX driver - set to 9 in-lbs (1.02 N.m)
T-20 TORX driver - set to 21 in-lbs (2.38 N.m)
5/16-in (8 mm) nutsetter or open end torque wrench - set to
10 in-lbs (1.13 N.m)
Use a 5/16-in torque wrench set to 10 in-lbs on all cable connections
except the front panel and rear panel bulkhead connectors. Torque these
connections to 21 in-lb.
About Installing the Upgrade
Products affected
Installation to be performed by
Estimated installation time
Estimated adjustment time
Estimated full instrument
calibration time
8
1
.
N5231A/B, N5232A/B, and N5239A/B Option 200
Keysight service center or personnel qualified by
Keysight
2 hours
0.5 hours
4.5 hours
Keysight Part Number
9300-1367
9300-0980
9300-0797
9300-1308
Qty
Part Number
1
N/A
1
N/A
1
N/A
Installation Note N5232-90101