Download Print this page

Keysight N5231A Installation Note page 16

Configurable test set and source attenuators upgrade kit option 200 to option 216

Advertisement

Installation Procedure for the Upgrade
Figure 3
Location of Cable Clamps to Secure W19, W22, and W23
— W19 (N5232-20025) A21 MASSY to REF 2 SOURCE OUT
— W23 (N5232-20024) A24 mixer brick to REF 2 RCVR R2 IN
— W24 (N5232-20023) A24 mixer brick to PORT 2 RCVR B IN
— W20 (N5232-20022) A26 test port 2 bridge coupler (arm) to PORT 2 CPLR
— W1 (reuse) (top-side cable) (N5235-20040) A4 13.5 GHz Source
16
* As shown in
Figure
3, install two cable clamps (part number 1400-1334)
to secure W19 (part number N5232-20025).
* As shown in
Figure
3, install three cable clamps (part number 1400-1334)
to secure W23 (part number N5232-20024).
ARM
Synthesizer board to test set deck cable bracket
Installation Note N5232-90101

Advertisement

loading

This manual is also suitable for:

N5231bN5232aN5232bN5239aN5239b