Download Print this page

Keysight N5231A Installation Note page 23

Configurable test set and source attenuators upgrade kit option 200 to option 216

Advertisement

Installation Procedure for the Upgrade
Step 16. Remove Option 200 License
Procedure Requirements
— The analyzer must be powered up and operating to perform this procedure.
— The Network Analyzer program must not be running.
— A keyboard and mouse must be connected to the network analyzer.
A Model Option 200 License Removal Procedure
For B models, refer to
1. To start the option enable utility, press UTILITY
2. Click the arrow in the Select Desired Option box. A list of available options
3. In the Select Desired Option list, click 200.
4. Click Remove.
B Model Option 200 License Removal Procedure
For A models, refer to
1. To start the Keysight License Manager, press Start > Keysight License
2. Right click on the desired option.
3. In the Select Desired Option list, click 200.
4. Click Remove.
5. A message displays stating that the option removal was successful.
Step 17. Enable Options 216
Procedure Requirements
— The analyzer must be powered up and operating to perform this procedure.
— The Network Analyzer program must not be running.
— Refer to the license message you received from Keysight: Verify that the
— A keyboard and mouse must be connected to the network analyzer.
Installation Note N5232-90101
"B Model Option 200 License Removal Procedure."
Option Enable
then
. An option enable dialog box will appear.
will appear.
"A Model Option 200 License Removal Procedure."
Manager > Keysight License Manager . A Keysight License Manager dialog
box will appear.
analyzer's model and serial numbers match those on the license message
you received from Keysight.
Service
System
, then
,
23

Advertisement

loading

This manual is also suitable for:

N5231bN5232aN5232bN5239aN5239b