Tools Required for the Installation

T-10 TORX driver - set to 9 in-lbs (1.02 N.m)
T-20 TORX driver - set to 21 in-lbs (2.38 N.m)
5/16-in (8 mm) nutsetter or open end torque wrench - set to 10 in-lbs (1.13 N.m)
Use a 5/16-in torque wrench set to 10 in-lbs on all cable connections except the front panel
CAUTION
and rear panel bulkhead connectors. Torque these connections to 21 in-lb.
About Installing the Upgrade
Products affected . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . N5231A, N5232A, and N5239A Option 200
Installation to be performed by . . . . . . . . . . . . . . . . . . . . . . Keysight service center or personnel qualified by
Estimated installation time . . . . . . . . . . . . . . . . . . . . . . . . . 2 hours
Estimated adjustment time . . . . . . . . . . . . . . . . . . . . . . . . . 0.5 hour
Estimated full instrument calibration time . . . . . . . . . . . . . 4.5 hours
6
Description
Keysight
Qty
Part Number
1
N/A
1
N/A
1
N/A