Download Print this page

Keysight N5231A Installation Note page 15

Configurable test set and source attenuators upgrade kit option 200 to option 216

Advertisement

Installation Procedure for the Upgrade
Step 8. Install the Cables
Install the Semirigid Cables
To see an image showing the location of these cables, click the Chapter 6
bookmarks "Bottom RF Cables, 2-Port Configuration, Option 216" in the PDF
Service Guide
Install the following new cables in the order listed.
— W22 (N5232-20031) A24 mixer brick to REF 1 RCVR R1 IN
Installation Note N5232-90101
Follow instructions carefully when making cable connections, especially
wire harness connections. Incorrect connections can destroy components,
resulting in additional customer costs.
Be careful not to damage the center pins of the semirigid cables. Some
flexing of the cables may be necessary but do not over-bend them.
Use a 5/16-in torque wrench set to 10 in-lbs on all cable connections
except the front and rear panel bulkhead connectors. On these, use a 9
mm nutsetter or open end torque wrench set to 21 in-lb.
1
. New parts are listed in
* As shown in
Figure
to secure W22 (part number N5232-20031).
1. See
"Downloading the Online PNA Service Guide" on page
Table 1 on page
3, install three cable clamps (part number 1400-1334)
9.
7.
15