Not" Operation - Rigol DS1000Z-E Series User Manual

Digital
Hide thumbs Also See for DS1000Z-E Series:
Table of Contents

Advertisement

RIGOL
Note:
operation.
Use HORIZONTAL
used to adjust the horizontal position and scale of the operation results.

"NOT" Operation

Perform logic "NOT" operation on the waveform voltage values of the specified
sources point by point and display the results. In operation, when the voltage value
of the source channel is greater than the threshold of the corresponding channel, it is
regarded as logic "1"; otherwise logic "0".
The results of logic NOT operation of one binary bit are as follows.
Table 6-5 Logic "NOT" Operation
A
0
1
Press MATH  Math  Operator to select "!A":
Press Operation to enable or disable the "NOT" operation function.
Press SourceA to select the desired channels (CH1 or CH2).
If source A selects one of the channels (CH1-CH2), press Thre.A and use
set the threshold of source A in logic operation.
Press Offset and use
Press Scale and use
Press Scale Reset to adjust the vertical scale of the operation results to the
optimal value according to the current configuration.
Press Options to set the start and end points of the operation results, enable or
disable waveform invert, etc.
― Press Start and use
― Press End and use
― Press Invert to enable or disable the inverted display function of the
― Press Sens. and use
― Press Auto Scale to enable or disable the auto scale function. When auto
6-14
Find Quality Products Online at:
scale is enabled, the instrument will adjust the vertical scale of the
operation results to the optimal value according to the current
configuration.
Smooth is grayed out and disabled. It is only available for differential
POSITION and HORIZONTAL
!A
1
0
to adjust the vertical position of the operation results.
to adjust the vertical scale of the operation results.
to set the end point of the operation results.
waveform.
from the analog signal on the source. The setting range is from 0 Div to
0.96 Div.
scale is enabled, the instrument will adjust the vertical scale of the
GlobalTestSupply
www.
Chapter 6 MATH and Measurement
to set the start point of the operation results.
to set the sensitivity of the digital signal converted
SCALE can also be
DS1000Z-E User Guide
.com