Load type
Placement
On trays allowing adequate space in-between
Pouched items
bags. Ensure that packs do not touch the walls of
the chamber. Place sterilization pouched items
with the paper side facing up.
Empty containers or non-
Upside down to prevent accumulation of water
perforated trays
Items made from different
On separate trays or wrapped/pouched
materials (stainless steel,
carbon steel, aluminum, etc.)
Instruments manufactured
Place paper among them and the trays to avoid
from carbon steel
rusty spots
LISA | Instructions for Use | VA131 2019 ENG Rev06 | 16/04/2020 | © 2020 W&H Sterilization Srl
PARTIAL LOAD
If the chamber is just partially loaded, place the load in such a way
that the space in-between the trays is maximized. Spread items
evenly on multiple trays. Below is a example with three trays.
Prepare the sterilizer
WARNINGS
Notice: Use only distilled or demineralized water (see "Feed water
specifications (EN 13060)" on page 126 for technical
requirements). Do not add any chemical / additive to the water.
Sterilization cycles
59