Diagnostics
Code
3xx
Check the door gasket. Clean or replace it if necessary.
Clean the chamber face side.
Clean the chamber filter.
Check if chamber filter is properly locked in the cap.
Check the load does not exceed the MAXIMUM WEIGHT LIMITS.
4xx
Clean water error (bad quality or low water level). Drain and refill the clean water tank.
5xx
Check if there are hurdles on the door locking area (chamber rack, loads, objects, ...).
Check the door gasket (wrong placed).
Check if the door can move freely without touching the trays or the load when closing.
Switch the sterilizer OFF and ON.
990
The cycle has been aborted by the user.
MESSAGES AND ALERTS
Notice: For any error not listed in this table, call technical service.
Message/Alert
Fill clean water tank.
Drain used water tank.
Please close the door.
102
Description
Description
There is not enough water in the tank to perform a
cycle.
The used water tank is full.
The door must be locked, but you didn't close it.
Repeat the cycle.
If the problem persists, call service.
Repeat the cycle.
If the problem persists, call service.
Repeat the cycle.
If the problem persists, call service.
Re-process the load.
Fill the water tank as requested.
Drain the water tank as requested.