PURPOSE OF THE TEST
The test is used to validate the sterilizer performance for textile load sterilization.
DESCRIPTION
It consists of several sheets of paper wrapped in a small packet with a chemical heat-sensitive indicator sheet in the middle. The colour
assumed by this indicator sheet at the end of the sterilization cycle gives the result of the test.
CARRY OUT THE TEST (IN ACCORDANCE WITH EN13060)
Empty the sterilizer chamber to ensure no load is present. Remove all trays from
1
the sterilization chamber, except the lowest one.
Place the Bowie and Dick test pack in the center of a tray in the lowest rack
2
position and close the chamber door.