Sterilizer tests

Helix test

CAUTION! Follow local/national guidelines on the frequency of testing.
PURPOSE OF THE TEST
The test is used to validate the sterilizer performance for hollow items.
DESCRIPTION
It consists of a 1.5 m long tube open on one side and closed with a capsule containing a chemical indicator strip on the other side. The colour
assumed by this indicator strip at the end of the sterilization cycle gives the result of the test.
CARRY OUT THE TEST (IN ACCORDANCE WITH EN13060)
Empty the sterilizer chamber to ensure no load is present. Remove all trays from
1
the sterilization chamber, except the lowest one.
Unscrew the tube capsule and place inside it an indicator strip according to the
2
instructions of the test manufacturer.
Screw the capsule.
3
52
LISA  | Instructions for Use |  VA131 2019 ENG Rev06 |  16/04/2020 |  © 2020 W&H Sterilization Srl