Sterilizer tests
WHAT TO DO WHEN THE TEST FAILED
Check, clean or replace the door gasket.
1
Clean the chamber face side and the chamber filter.
2
Repeat the Vacuum test. See "Carry out the test" on the previous page.
3
If the test fails repeatedly, call technical service.
4