800-cycle maintenance

REPLACING THE DOOR GASKET
A video animation is available on https://video.wh.com/en_global/dental-faq.
Notice: If you replace this consumable before the maintenance due date you have to reset the cycle counter. See "Replace the consumable
before the maintenance due date" on page  79.
Proceed as follows:
Fully open the chamber door.
1
Remove the used door gasket by hand.
2
Carefully clean the seal seat and the chamber face side with a damp, lint-free
3
cloth.
Moisten the new seal with water to facilitate its placement.
4
LISA  | Instructions for Use |  VA131 2019 ENG Rev06 |  16/04/2020 |  © 2020 W&H Sterilization Srl
Maintenance
95