MAXIMUM LOAD FOR INSTRUMENTS
Note: The load given includes the trays, the containers and everything is put into the chamber, with the sole exclusion of the tray rack.
LISA 17
B Universal 134
4.5 kg (10 lbs)
B Extended 134
4.5 kg (10 lbs)
B Universal 121
4.5 kg (10 lbs)
S Fast 134
MAXIMUM LOAD FOR CONTAINERS
The correct dryness can only be obtained with the ECO DRY mode.
LISA 17
B Universal 134 9.0 kg (19.8 lbs) 9.0 kg (19.8 lbs)
B Extended 134 9.0 kg (19.8 lbs) 9.0 kg (19.8 lbs)
B Universal 121 9.0 kg (19.8 lbs) 9.0 kg (19.8 lbs)
S Fast 134
not suitable
LISA  | Instructions for Use |  VA131 2019 ENG Rev06 |  16/04/2020 |  © 2020 W&H Sterilization Srl
Bagged
LISA 22
6.0 kg (13.2 lbs)
6.0 kg (13.2 lbs)
6.0 kg (13.2 lbs)
-
-
LISA 22
not suitable
Instruments
Unwrapped
LISA 17
LISA 22
4.5 kg (10 lbs)
6.0 kg (13.2 lbs)
4.5 kg (10 lbs)
6.0 kg (13.2 lbs)
4.5 kg (10 lbs)
6.0 kg (13.2 lbs)
2.0 kg (4.4 lbs)
2.0 kg (4.4 lbs)
Technical data
Porous
LISA 17
LISA 22
1.5 kg (3.3 lbs)
2.0 kg (4.4 lbs)
1.5 kg (3.3 lbs)
2.0 kg (4.4 lbs)
1.5 kg (3.3 lbs)
2.0 kg (4.4 lbs)
-
-