Prepare The Sterilizer - W&H Lyla Instructions For Use Manual

Table of Contents

Advertisement

Load type
Placement
On trays allowing adequate space in-between
Pouched items
bags. Ensure that packs do not touch the walls of
the chamber. Place sterilization pouched items
with the paper side facing up.
Empty containers or non-
Upside down to prevent accumulation of water
perforated trays
Items made from different
On separate trays or wrapped/pouched
materials (stainless steel,
carbon steel, aluminium, etc.)
Instruments manufactured
Place paper among them and the trays to avoid
from carbon steel
rusty spots
Lyla  | Instructions for Use |  RIN-210 ENG Rev02 |  15/09/2022 |  © W&H Sterilization Srl
PARTIAL LOAD
If the chamber is just partially loaded, place the load in such a way
that the space in-between the trays is maximized. Spread items
evenly on multiple trays. Below is a example with three trays.

Prepare the sterilizer

WARNINGS
Notice: use only distilled or demineralized water (see "Feed water
specifications (EN 13060)" on page  109 for technical requirements).
Do not add any chemical / additive to the water.
Sterilization cycles
49

Advertisement

Table of Contents
loading

This manual is also suitable for:

Rin-210-17Rin-210-22

Table of Contents