Maintenance; Warnings For Maintenance Operations; Ordinary Maintenance - W&H Lisa VA131-17 Instructions For Use Manual

Hide thumbs Also See for Lisa VA131-17:
Table of Contents

Advertisement

Maintenance

CONTENTS
This section deals with the following subjects:

Warnings for maintenance operations

Ordinary maintenance

Monthly or 50-cycle maintenance
400-cycle maintenance
800-cycle or biannual maintenance
800-cycle maintenance
4000 cycle or five-year maintenance
Extraordinary maintenance
Disposal
Warnings for maintenance operations
WARNINGS
WARNING! Turn the sterilizer OFF and remove the power
cord before beginning any maintenance. Follow all health,
safety, cross-infection and cross-contamination protocols.
Maintenance operation shall be done at illumination level of
215 lx (±15 lx) to 1500 lx (±15 lx).
Before making any operation, ward off unauthorized
personnel from the working area.
LISA  | Instructions for Use |  VA131 2019 ENG Rev06 |  16/04/2020 |  © 2020 W&H Sterilization Srl
CAUTION! Before accessing the chamber and the connected
parts, be sure that the sterilizer is cold.
Notice: Follow the instructions in this chapter when carrying out any
77
maintenance on the sterilizer.
77
80
85
Ordinary maintenance
88
95
97
MAINTENANCE BY THE USER
98
1
Frequency
99
Monthly
6 month
1
Cycles
Operation
Cleaning the door gasket and the chamber face
50
side. See "Cleaning the door gasket and the
chamber face side" on page  80
Clean the chamber, trays and the rack. See
"Cleaning the chamber and the chamber
accessories" on page  81
Cleaning the chamber filter. See "Cleaning the
chamber filter" on page  82
Cleaning the external surfaces of the sterilizer.
See "Cleaning the external surfaces of the
sterilizer" on page  84
800
Clean both water tanks. See "800-cycle or
biannual maintenance" on page  88.
77

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Lisa va131-22

Table of Contents