Typical Loads; Prepare The Sterilizer - W&H Lexa MN-111 Instructions For Use Manual

Hide thumbs Also See for Lexa MN-111:
Table of Contents

Advertisement

Sterilization cycles

Typical loads

EXAMPLES
An example of a typical load used during the tests performed by
W&H is reported below:
Program
Load example
Pouches Medium Load
Pouches Large Load
Pouches & Wrapped Cassettes
Wrapped Cassettes
48

Prepare the sterilizer

WARNINGS
Notice: use only distilled or demineralized water (see "Water quality"
on page  95 for technical requirements). Do not add any chemical /
additive to the water.
FILLING THE CLEAN WATER TANK
If no water system is connected you have to fill the tank:
Switch the sterilizer ON and open the chamber door.
1
Fill the clean water tank with distilled or demineralized water
2
until the sterilizer makes a sound. See "Technical data" on
page  92  for the tank volume.
INSERTING THE CHAMBER RACK INTO THE STERILIZER
CAUTION! Risk of burns. Before touching the chamber rack or
contents, ensure the sterilization chamber is not hot.
Open the chamber door and align the chamber rack at the
1
center/bottom of the chamber.
Push the chamber rack gently into position until it clicks into
2
place.
Insert cassettes horizontally or vertically, or insert trays. See
3
"Load maintenance and preparation" on page  45 for load
requirements .
Close the door.
4
Lexa  | Instructions for Use |  MN-111 ENG Rev14 |  03/04/2023 |  © W&H Sterilization Srl