KaVo ESTETICA E50 Life Instructions For Use Manual

KaVo ESTETICA E50 Life Instructions For Use Manual

Hide thumbs Also See for ESTETICA E50 Life:
Table of Contents

Advertisement

Instructions for use
ESTETICA E50 Life

Advertisement

Table of Contents
loading

Summary of Contents for KaVo ESTETICA E50 Life

  • Page 1 Instructions for use ESTETICA E50 Life...
  • Page 2 Distributed by: Manufacturer: KaVo Dental Technologies, LLC Kaltenbach & Voigt GmbH 11727 Fruehauf Drive Bismarckring 39 Charlotte, NC 28273 USA D-88400 Biberach Tel. 847-550-6800 www.kavo.com Fax 847-550-6825...
  • Page 3: Table Of Contents

    Instructions for use ESTETICA E50 Life Table of contents Table of contents 1 User instructions ........................ 7 1.1 User guide ........................7 1.1.1 Abbreviations ..................... 7 1.1.2 Symbols ......................7 1.1.3 Target group ...................... 7 1.2 Service ......................... 8 1.3 Terms and conditions of warranty..................8 1.4 Transportation and storage.....................
  • Page 4 Instructions for use ESTETICA E50 Life Table of contents 4 Operation..........................47 4.1 Switching the device on and off..................47 4.2 Adjusting the dental chair ....................48 4.2.1 Adjusting the arm rest (optional)................48 4.2.2 Adjust head rest ....................49 4.2.3...
  • Page 5 Instructions for use ESTETICA E50 Life Table of contents 4.10.3 Using the three function handpiece............... 99 4.10.4 Using the multifunctional handpiece..............101 4.10.5 Using the PiezoLED ..................... 104 4.11Using the KL 703 LED / KL 702 in ENDO mode (optional accessory) ........105 4.11.1 General information ....................
  • Page 6 Instructions for use ESTETICA E50 Life Table of contents 9 Troubleshooting ......................... 151 10Data on electromagnetic compatibility according to EN 60601-1-2 ........156 10.1Electromagnetic Transmissions..................156 10.2Resistance to electromagnetic interference ............... 156 10.3Recommended safe distance between portable and mobile HF telecommunications equipment and the treatment unit ...................
  • Page 7: User Instructions

    Instructions for use ESTETICA E50 Life 1 User instructions | 1.1 User guide 1 User instructions 1.1 User guide Requirement Read these instructions prior to first startup to avoid misuse and prevent dam- age. 1.1.1 Abbreviations Explanation brevi- ation Instructions for use...
  • Page 8: Service

    For further information, please visit: www.kavo.com 1.3 Terms and conditions of warranty KaVo provides the final customer with a warranty that the product cited in the handover certificate will function properly and guarantees zero defects in the material or processing for a period of 12 months from data of purchase, subject...
  • Page 9: Damage In Transit

    Terms and Conditions, Art. 28). Outside Germany Note KaVo shall not be held liable for damage arising from transportation. The shipment must be checked on arrival. If the packaging is visibly damaged on delivery, please proceed as follows: 1. The recipient of the package must record the loss or damage on the delivery receipt.
  • Page 10: Information On The Packaging: Storage And Transportation

    Instructions for use ESTETICA E50 Life 1 User instructions | 1.4 Transportation and storage 1.4.3 Information on the packaging: Storage and transportation Note Please keep the packaging in case you need to return the product for servi- cing or repair.
  • Page 11: Safety

    Instructions for use ESTETICA E50 Life 2 Safety | 2.1 Description of safety instructions 2 Safety 2.1 Description of safety instructions 2.1.1 Warning symbol Warning symbol 2.1.2 Structure DANGER The introduction describes the type and source of the hazard. This section describes potential consequences of non-compliance.
  • Page 12: General

    When connecting IT equipment to the the medial electrical system, observe EN 60601-1. Note Charge the wireless foot control with the charger supplied by KaVo only. Note The wireless foot control charger may only be used indoors and must be pro- tected from moisture.
  • Page 13 ISO 7494 equipped with a patient chair in accordance with ISO 6875. This KaVo product is designed for use in dentistry only and may only be used by trained medical personnel. Any other type of use is not permitted.
  • Page 14 Note The MULTIflex couplings, the current K/KL motors, and the ultrasonic scaler hoses of KaVo are equipped as standard with a protective device to prevent treatment water from being drawn back into the treatment centre via the handpieces. If products from other manufacturers are used at the standard-...
  • Page 15: Safety Instructions

    DANGER Explosion hazard. Risk of fatal injury. ▶ Do not use KaVo product in areas subject an explosion hazard. WARNING Inappropriate operating conditions. Impairment of the electrical safety of the device. ▶ It is essential to comply with the operating conditions specified in the "Technical Specifications"...
  • Page 16: Product-Specific

    Instructions for use ESTETICA E50 Life 2 Safety | 2.3 Safety instructions NOTICE Premature wear and malfunctions from improper servicing and care. Reduced product life. ▶ Perform regular proper care and servicing! CAUTION Risks from electromagnetic fields. Electromagnetic fields might interfere with the functions of implanted systems (such as pacemakers).
  • Page 17 Instructions for use ESTETICA E50 Life 2 Safety | 2.3 Safety instructions CAUTION Sitting down on a dental chair that is in horizontal orientation is asso- ciated with a risk of injury. ▶ Do not sit on the head or foot end of the patient chair when it is in a hori- zontal position.
  • Page 18 Instructions for use ESTETICA E50 Life 2 Safety | 2.3 Safety instructions CAUTION Health damage due to germ formation. Infection hazard. ▶ Before starting, rinse all the water drain lines with no handpieces attached. ▶ Before start-up and after downtimes (weekends, holidays, vacations, etc.), rinse or purge the air and water lines with air.
  • Page 19: Product Description

    Instructions for use ESTETICA E50 Life 3 Product description | 3.1 Treatment unit versions 3 Product description The ESTETICA Dental Treatment Unit and Accessories serves as a base that in- cludes components to deliver air, water, electrical power, and vacuum to dental handpieces, instruments, and accessories.
  • Page 20: Kavo Estetica E50 Life S

    Instructions for use ESTETICA E50 Life 3 Product description | 3.1 Treatment unit versions 3.1.2 KaVo ESTETICA E50 Life S 20 / 162...
  • Page 21: Patient Chair Standard And Compactchair

    Instructions for use ESTETICA E50 Life 3 Product description | 3.2 Patient chair Standard and COMPACTchair 3.2 Patient chair Standard and COMPACTchair ① Headrest ② Backrest ③ Chair base ④ Seat ⑤ Arm rest (optional) 21 / 162...
  • Page 22: Device Body With Patient Element (Optional)

    Instructions for use ESTETICA E50 Life 3 Product description | 3.3 Device body with patient element (optional) 3.3 Device body with patient element (optional) ① Patient element (optional) ② Unit body The central control is housed in the device body.
  • Page 23 Instructions for use ESTETICA E50 Life 3 Product description | 3.3 Device body with patient element (optional) E50 Life with Centramat ① Patient element (optional) ② Unit body The central control is housed in the device body. ③ Centramat ④ Spittoon bowl (supplementary equipment) ⑤...
  • Page 24: Dentist Unit Versions

    Instructions for use ESTETICA E50 Life 3 Product description | 3.4 Dentist unit versions 3.4 Dentist unit versions 3.4.1 TM table ① Handle ② Three-function handpiece or multi- functional handpiece ③ Turbine (multiflex coupling) ④ INTRA LUX motor KL 703 or INTRA LUX motor KL 701 ⑤...
  • Page 25: S Table

    Instructions for use ESTETICA E50 Life 3 Product description | 3.4 Dentist unit versions 3.4.2 S table Note The holder assignment and arrangement of the instruments can be changed as needed and does not have to follow the picture. ① Site for affixing the USB interface ②...
  • Page 26: Assistant Element - Versions

    Instructions for use ESTETICA E50 Life 3 Product description | 3.5 Assistant element – Versions 3.5 Assistant element – Versions 3.5.1 Standard assistant unit ① Three-function or multifunctional ② Spray mist suction handpiece ③ Control element ④ Saliva ejector ⑤ Satelec Mini LED ⑥...
  • Page 27: Assistant Element Right, Left (Optional)

    Instructions for use ESTETICA E50 Life 3 Product description | 3.5 Assistant element – Versions 3.5.2 Assistant element right, left (optional) Assistant element right, left with height adjustment (optional) ① Three-function handpiece ② Spray mist suction ③ Control element ④ Saliva ejector ⑤...
  • Page 28: Three Function Handpiece (3F Handpiece)

    Instructions for use ESTETICA E50 Life 3 Product description | 3.6 Three function handpiece (3F handpiece) 3.6 Three function handpiece (3F handpiece) ① MF handpiece hose ② Gripping sleeve ③ Media buttons (air/water) ④ Labelled blue: Three-function hand- piece (3F handpiece) ⑤...
  • Page 29: X-Ray Viewer 1440

    3 Product description | 3.8 X-ray viewer 1440 3.8 X-ray viewer 1440 X-ray viewer 1440 Note The KaVo X-ray viewer 1440 is a type 1 radiological viewing device in accord- ance as defined in DIN 6856-3. 3.9 Controls 3.9.1 Dentist's unit TM table...
  • Page 30: Dentist's Unit S Table

    Instructions for use ESTETICA E50 Life 3 Product description | 3.9 Controls 3.9.2 Dentist's unit S table Group of keys for menu selection Group of keys for illumination (MEMOdent-menu) Group of keys for hygiene Group of keys for the timer Group of keys for the dental chair 3.9.3 Assistant unit...
  • Page 31 Instructions for use ESTETICA E50 Life 3 Product description | 3.9 Controls Assistant element key Dentist element key Name "Chair up" key "AP 0" key (automatic position 0) "Chair down" key "SP" key (rinsing position) "LP" key (last position) "AP" key...
  • Page 32 Instructions for use ESTETICA E50 Life 3 Product description | 3.9 Controls Group of keys for hygiene Name Control element "Tumbler filler" key Dentist element and as- sistant element "Bowl rinsing" key Dentist element and assistant element "Bell" key dentist element only "Intensive germ reduc-...
  • Page 33 Instructions for use ESTETICA E50 Life 3 Product description | 3.9 Controls Mat. no. Control element "Timer 4" key Dentist element only 33 / 162...
  • Page 34: Foot Control

    Instructions for use ESTETICA E50 Life 3 Product description | 3.9 Controls 3.9.5 Foot control The footswitches of the foot control have two functions. The functions of the footswitches depend on if an instrument is mounted or removed. Foot control Premium and wireless foot control | Foot control Standard...
  • Page 35: Rating Plate And Identification Plate

    Instructions for use ESTETICA E50 Life 3 Product description | 3.10 Rating plate and identification plate 3.10 Rating plate and identification plate Rating plate Internal attachment site for rating plate 400s Outside attachment site for rating plate Year and month manufactured - Serial number...
  • Page 36 Instructions for use ESTETICA E50 Life 3 Product description | 3.10 Rating plate and identification plate Type B applied part Operating mode: Operating time of the patient chair: 25 seconds Pause time of the patient chair: 400 seconds (The permissible operating times correspond to common dental pro- cedure.)
  • Page 37 Instructions for use ESTETICA E50 Life 3 Product description | 3.10 Rating plate and identification plate Rating plate and dentist element ID Site of attachment of nameplate and type BF applied parts ID on dentist element Miscellaneous signs Attachment site back-side of wireless foot control...
  • Page 38: Rating Plate 1440

    Instructions for use ESTETICA E50 Life 3 Product description | 3.10 Rating plate and identification plate Serial number CE mark according to 93/42/EEC medical devices Sterilisable up to 135 For disposal information, see also: Purpose – Intended use Follow the instructions for use 3.10.1 Rating plate 1440...
  • Page 39: Technical Data

    Instructions for use ESTETICA E50 Life 3 Product description | 3.11 Technical data 3.11 Technical data Drilling template and setup plan Installation plan (Mat. no. 2 sheets each right-handed and 2 3.002.4533) sheets left-handed Installation plan with COMPACTchair 2 sheets each right-handed and 2 (Mat.
  • Page 40 Instructions for use ESTETICA E50 Life 3 Product description | 3.11 Technical data Wireless foot control RF Technology Proprietary 2,4 GHz ISM band Sys- Modulation GFSK Protocol KaVo specific RF Frequencies 2404MHz - 2478MHz (38 Channels) RF Power 0dBm (1mW) Range Indoor <...
  • Page 41 Instructions for use ESTETICA E50 Life 3 Product description | 3.11 Technical data Water pressure 1.5 ± 0.3 bar; Flow pressure; 4 x manometer Max. static pressure water 2.5 ± 0.3 bar Water flow 80 ± 10 ml/min Air pressure 3.3 ±...
  • Page 42 Note In conjunction with the "DVGW water block with integrated water germ re- duction" a water germ reduction unit is installed in dental units from KaVo. The germ reduction liquid OXYGENAL 6 cleaner is continually added to the water at a concentration which is harmless for persons, but hygienically ef- fective to maintain the quality of the treatment water by making it more diffi- cult for microorganisms to multiply and spread.
  • Page 43 Instructions for use ESTETICA E50 Life 3 Product description | 3.11 Technical data Free drainage according to DIN EN Water block DVGW, water bottle 1717 - DVGW certified DVGW, register no.: AS-0630BT0111 Water quality Tap water, cold water connection Water hardness 1.5 to 2.14 mmol/l ≙...
  • Page 44 Instructions for use ESTETICA E50 Life 3 Product description | 3.11 Technical data Suction Pressure drop at the connecting point ① Semi-dry and wet suction ② Dry suction Note If the negative static pressure is > 180 mbar, the unit must be equipped with the negative pressure regulating valve assembly kit.
  • Page 45 Instructions for use ESTETICA E50 Life 3 Product description | 3.11 Technical data Operating environment WARNING Inappropriate operating conditions. Impairment of the electrical safety of the device. ▶ It is essential to comply with the operating conditions specified in the "Technical Specifications"...
  • Page 46 Instructions for use ESTETICA E50 Life 3 Product description | 3.11 Technical data Weight Treatment centre (TM) with 210 kg gross, 160 kg net patient chair Standard Includes steel mounting plate and 320 kg gross, 266 kg net patient communication...
  • Page 47: Operation

    Instructions for use ESTETICA E50 Life 4 Operation | 4.1 Switching the device on and off 4 Operation 4.1 Switching the device on and off Note Always switch the machine off before leaving the office. E50 Life without/with DEKAmat/CENTRAmat ▶ Switch on the device using the main switch.
  • Page 48: Adjusting The Dental Chair

    Direct contact between the skin of the patient and the cushion during patient positioning To improve hygiene and increase the service life, KaVo recommends to use ap- propriate covers (e.g. head protection bags) to protect the cushions that are regularly in contact with the skin of the patient. Sweat or hair and skin care products can be a serious strain on the cushions.
  • Page 49: Adjust Head Rest

    Instructions for use ESTETICA E50 Life 4 Operation | 4.2 Adjusting the dental chair 4.2.2 Adjust head rest Setting locking dial of 2-joint headrest (standard) CAUTION Adjusting the headrest. Injury to the neck muscles. ▶ Make sure that the patient is aware of the headrest being adjusted.
  • Page 50: Positioning The Dental Chair Manually

    Instructions for use ESTETICA E50 Life 4 Operation | 4.2 Adjusting the dental chair Setting locking dial of 2-joint headrest (optional) CAUTION Adjusting the headrest. Injury to the neck muscles. ▶ Make sure that the patient is aware of the headrest being adjusted.
  • Page 51 Instructions for use ESTETICA E50 Life 4 Operation | 4.2 Adjusting the dental chair Positioning the chair and backrest manually using the dentist or assistant unit Use the following buttons to adjust the chair height and position of the back-...
  • Page 52: Automatic Positioning Of Dental Chair

    Instructions for use ESTETICA E50 Life 4 Operation | 4.2 Adjusting the dental chair 4.2.4 Automatic positioning of dental chair CAUTION Danger of injury from overload or dynamic load. Patient chair may be damaged by overloading it. ▶ Do not subject the patient chair to a load exceeding its limit (Standard pa- tient chair 185 kg/ COMPACTchair patient chair 135 kg).
  • Page 53 Instructions for use ESTETICA E50 Life 4 Operation | 4.2 Adjusting the dental chair It is for example recommendable to save the sitting down and getting up posi- tion using the "AP 0" key and the rinsing position with the "SP" key.
  • Page 54 Instructions for use ESTETICA E50 Life 4 Operation | 4.2 Adjusting the dental chair "AP 1" button: treatment position, e.g. for lower jaw treatment "AP 2" button: treatment position, e.g. for upper jaw treatment "Collape position" button: collapse position ▶ Move the chair to the desired position.
  • Page 55 Instructions for use ESTETICA E50 Life 4 Operation | 4.2 Adjusting the dental chair ð LED of the pressed button lights up. The chair position is saved. Recalling automatic positions with the foot control Note If an instrument is removed, the chair functions of the foot control are blocked.
  • Page 56: Safety Shut-Off

    Instructions for use ESTETICA E50 Life 4 Operation | 4.2 Adjusting the dental chair Saving an automatic position with the foot control ① Spray preselection/AP footswitch ② Blown air/AP footswitch ③ Foot pedal The chair positions can be saved on two footswitches; the standard setting is as follows: ▪...
  • Page 57 Instructions for use ESTETICA E50 Life 4 Operation | 4.2 Adjusting the dental chair CAUTION Risk of injury when moving the patient or patient chair. The patient or treatment personnel can be pinched or crushed. ▶ Position all moving parts, such as dentist element, assistant element, oper- ating light, screens, etc., outside the collision range when you move the pa-...
  • Page 58 Instructions for use ESTETICA E50 Life 4 Operation | 4.2 Adjusting the dental chair Item Safety switch-off ac- LED on assistant LED on dentist ele- tuated element ment ④ Bracket on the foot control Base plate ⑤ ⑥ Bench Patient chair COMPACTchair Safety shutoff for the COMPACTchair patient chair ①...
  • Page 59 Instructions for use ESTETICA E50 Life 4 Operation | 4.2 Adjusting the dental chair The safety shutoff occurs if a movement angle has been exceeded or if a part of the treatment centre collides with an object. If a person or object actuates a safety shutoff, the chair immediately stops moving.
  • Page 60: Moving The Patient Chair

    Instructions for use ESTETICA E50 Life 4 Operation | 4.3 Moving the patient chair 4.3 Moving the patient chair Patient chair Standard Patient chair COMPACTchair 4.4 Moving the dentist unit NOTICE Damage from overloading the dentist element. Exceeding the maximum weight of more than 2 kg by adding handpieces, ac- cessories, etc., can cause damage.
  • Page 61: Moving The Dentist Unit Tm

    Instructions for use ESTETICA E50 Life 4 Operation | 4.4 Moving the dentist unit Note Do not pull the dentist unit by the instrument hose. ▶ To adjust the height of the dentist unit, release the brake, adjust the height, and reset the brake.
  • Page 62: Moving The Patient Element (Optional)

    Instructions for use ESTETICA E50 Life 4 Operation | 4.5 Moving the patient element (optional) 4.5 Moving the patient element (optional) 4.5.1 Swinging the patient unit by hand The swinging range is about 250 Note When the patient unit is swung over the patient chair, the safety shutoff is activated.
  • Page 63: Moving The Assistant Element Right, Left (Optional)

    Instructions for use ESTETICA E50 Life 4 Operation | 4.6 Moving the assistant element ▶ To set a higher level, pull the assistant unit upward gently until it audibly locks in place. ▶ To set a lower level, pull the assistant unit all the way up until the lock re- leases, and then lower the assistant unit.
  • Page 64 Instructions for use ESTETICA E50 Life 4 Operation | 4.6 Moving the assistant element Pivoting range of assistant element r, l (optional) ▶ Move the backrest up before swinging the assistant element. ▶ Move the assistant element to the desired position in its swinging range.
  • Page 65: Using Functions Through The Menu

    Instructions for use ESTETICA E50 Life 4 Operation | 4.7 Using functions through the menu ▶ Undo the clamping screw and push the assistant element into the desired position. ▶ Re-tighten the clamping screw. 4.7 Using functions through the menu 4.7.1 Using the user menu...
  • Page 66 Instructions for use ESTETICA E50 Life 4 Operation | 4.7 Using functions through the menu Starting and closing the user menu ▶ Press the "Next" key (S6) to switch to the user menu. ð Option 1 "Number of users" or most recently used option is displayed.
  • Page 67 Instructions for use ESTETICA E50 Life 4 Operation | 4.7 Using functions through the menu Option 2: Set tumbler filling time ▶ Press the "reduce value" or "increase value" key to select a tumbler filling time from 0 - 51 seconds.
  • Page 68 Instructions for use ESTETICA E50 Life 4 Operation | 4.7 Using functions through the menu Option 8: Set the date ▶ Press the "Increase value" or "Decrease value" keys to set the date. ▶ Press the"Move cursor" (S3) key to switch between day, month, and year.
  • Page 69 Instructions for use ESTETICA E50 Life 4 Operation | 4.7 Using functions through the menu Option 11: Setting the menu language ▶ Press the "increase value" or "reduce value" keys to set the desired lan- guage. Option 12: Setting dimming mode for the LED lamp.
  • Page 70: Standby Menu

    Instructions for use ESTETICA E50 Life 4 Operation | 4.7 Using functions through the menu ▶ Press the "Increase value" or "Reduce value" buttons keys to set the colour temperature for the LED lamp. The factory setting is 5500 K.
  • Page 71 Instructions for use ESTETICA E50 Life 4 Operation | 4.7 Using functions through the menu Example standby menu with selection buttons Permitting level switching Level switching is deactivated in the basic state. The level switching symbol displays the current dentist.
  • Page 72 Instructions for use ESTETICA E50 Life 4 Operation | 4.7 Using functions through the menu Selecting the dentist Dentist selection level when switching is deactivated ▶ Press the “Dentist” key. ð The display changes to the settings menu. ▶ Press the key for the desired dentist (dentist 1 to dentist 6).
  • Page 73: Using The Memodent Menu

    Instructions for use ESTETICA E50 Life 4 Operation | 4.7 Using functions through the menu ▶ Press the "OK" selection key ③ to exit from the display of status messages. Error messages in the status display See also: 2 9 Troubleshooting, Page 151 4.7.3 Using the MEMOdent menu...
  • Page 74: Comfortdrive

    Instructions for use ESTETICA E50 Life 4 Operation | 4.7 Using functions through the menu Setting the cooling level ▶ Adjust the cooling with the selection button for "Cooling status". Symbol Feature No cooling Spray air cooling status Spray cooling status...
  • Page 75: Changing Piezoled Settings In The Memodent Menu

    Instructions for use ESTETICA E50 Life 4 Operation | 4.7 Using functions through the menu ▶ Press "Preselect level" button for 4 seconds in order to change settings. ð The display switches to the motor settings menu. ▶ Press the "Save" key to save the values. You can save after setting each value, or after setting all values.
  • Page 76 Instructions for use ESTETICA E50 Life 4 Operation | 4.7 Using functions through the menu ▶ Press "Preselect level" button for 4 seconds. ð The display switches to the settings menu of the PiezoLED. ▶ Press the "Save" key to save the values. You can save after setting each value, or after setting all values.
  • Page 77: Changing Settings For The Multifunctional Handpiece In The Memodent Menu

    Instructions for use ESTETICA E50 Life 4 Operation | 4.7 Using functions through the menu ▶ For the amount of spray water for each tip, please refer to the PiezoLED In- structions for Use. See also: 2 Instructions for Use PiezoLED ▶...
  • Page 78: Use The Timer

    Instructions for use ESTETICA E50 Life 4 Operation | 4.7 Using functions through the menu ▶ Press the "Save" key to save the values. You can save after setting each value, or after setting all values. ð Saving is acknowledged with a tone.
  • Page 79: Using The Conexiocom Menu

    The function of the CONEXIOcom menu is to control the display of previously recorded and saved images and videos. In order to use the function, the unit must have access to the data of the KaVo Software "CONEXIO" software. For details on the configuration, please refer to the "CONEXIO" installation instruc- tions.
  • Page 80 Instructions for use ESTETICA E50 Life 4 Operation | 4.7 Using functions through the menu Icon Setting Info line This line displays the active patient name (if selected in CONEXIO) un- der which the acquired data are stored. If no patient is selected, images and videos are stored directly un- der "unassigned patient".
  • Page 81: Using Function Through The Dentist Or Assistant Unit

    Instructions for use ESTETICA E50 Life 4 Operation | 4.8 Using function through the dentist or assistant unit 4.8 Using function through the dentist or assistant unit 4.8.1 Using the hygiene functions The following buttons are available for the hygiene functions: Feature The tumbler is being filled.
  • Page 82 KaVo KEY Laser III or KEY Laser 3+ can lead to the unintentional activation of the KaVo KEY Laser III and KEY Laser 3+. ▶ When using the KaVo KEY Laser III or the KEY Laser 3+, switch the operat- ing light to laser mode.
  • Page 83 KaVo DIAGNOdent can lead to faulty measurements. ▶ Switch the operating light to laser mode when using the KaVo DIAGNOdent. ▶ Or switch off the operating light, do not use KaVo DIAGNOdent and operat- ing light KaVoLUX 540 LED simultaneously. CAUTION Premature hardening of composite fillings.
  • Page 84 Instructions for use ESTETICA E50 Life 4 Operation | 4.8 Using function through the dentist or assistant unit ▶ Press the "Treatment light" button. ð ▶ Hold your hand just in front of the sensor. ð Treatment light is switched on, Display LED shining.
  • Page 85 Instructions for use ESTETICA E50 Life 4 Operation | 4.8 Using function through the dentist or assistant unit ▶ Hold your hand in front of the sensor for 2 sec. ð The COMPOsave mode switches on. ð The COMPOsave mode can be recognised by the yellow light.
  • Page 86 Therefore a colour comparison should not be carried out in laser mode. In laser mode, another light mode is generated, which has no negative influ- ence on the KaVo KEY Laser III, the KEY Laser 3+ or the KaVo DIAGNOdent. 86 / 162...
  • Page 87 Instructions for use ESTETICA E50 Life 4 Operation | 4.8 Using function through the dentist or assistant unit Sensor KaVoLUX 540 LED ▶ Press the "operating light“ button and the "operating light dimming" button simultaneously on the dentist element of the treatment unit.
  • Page 88: Using The Timer

    Instructions for use ESTETICA E50 Life 4 Operation | 4.8 Using function through the dentist or assistant unit Operation of the 3D joint ▶ Turn the switching ring to the left until it snaps into place. ð Treatment light can be rotated 45° to the left or 45° to the right.
  • Page 89: Operating The Foot Switch

    Instructions for use ESTETICA E50 Life 4 Operation | 4.9 Operating the foot switch Select the timer time ▶ To start a timer time, e.g. Timer 1, press the "Timer 1" button. ð Time on the timer starts to run. A beep is issued after the timer time is elapsed.
  • Page 90 Instructions for use ESTETICA E50 Life 4 Operation | 4.9 Operating the foot switch Item Labelling Function ① LED display Status display / charge status dis- play On/Off switch On/off switch to prevent deep dis- ② charge during long periods of non- use.
  • Page 91: Establishing A Connection Between The Wireless Foot Control And The Treatment Unit

    Instructions for use ESTETICA E50 Life 4 Operation | 4.9 Operating the foot switch 4.9.3 Establishing a connection between the wireless foot control and the treatment unit CAUTION Loss of functionality due to interruption of wireless connection Other coexistent wireless devices, which are working in the same frequency band, could disturb the wireless connection of the foot control.
  • Page 92 Instructions for use ESTETICA E50 Life 4 Operation | 4.9 Operating the foot switch A combination of keys must be entered on the wireless foot control in the proper sequence during a countdown of 30 seconds. ▶ Press the foot pedal, then move the cross switch toward "Chair up", and then actuate and hold the stirrup switch until OK appears on the display.
  • Page 93: Positioning The Patient Chair With The Foot Control

    Instructions for use ESTETICA E50 Life 4 Operation | 4.9 Operating the foot switch CAUTION Improper use of the wireless foot control Damage or malfunction ▶ In case of improper use (such as cleaning), turn off the wireless foot control or the treatment unit.
  • Page 94: Setting The Cooling Condition

    Instructions for use ESTETICA E50 Life 4 Operation | 4.9 Operating the foot switch CAUTION Centring in the middle is effected for the foot control by a positioning motor. If the positioning motor breaks down, switching from or into the middle posi- tion using the foot control is no longer feasible.
  • Page 95: Preselect Counterclockwise Motor Rotation

    Instructions for use ESTETICA E50 Life 4 Operation | 4.9 Operating the foot switch ▶ Press the "Blown air" foot button. ð As long as the footswitch is pressed, blown air exits the removed handpiece (not with the PiezoLED) 4.9.10 Preselect counterclockwise motor rotation ▶...
  • Page 96: Using Conexiocom (Fee-Based Additional Option)

    Instructions for use ESTETICA E50 Life 4 Operation | 4.9 Operating the foot switch 4.9.13 Using CONEXIOcom (fee-based additional option) Setting Stirrup switch ① Discard image/video Press briefly - deletes the selected image/video Previous image/video ② Select previous image/video ③...
  • Page 97: Using Instruments

    ▶ Charge the wireless foot control with the charger supplied only. Note Charge the wireless foot control with the charger supplied by KaVo only. Note The wireless foot control charger may only be used indoors and must be pro- tected from moisture.
  • Page 98 Instructions for use ESTETICA E50 Life 4 Operation | 4.10 Using instruments The suction flow of the saliva ejector or spray mist suction device can be re- duced or blocked with the slide valves integrated in the handpieces. ▶ Move the slide valve completely upward.
  • Page 99: Using The Three Function Handpiece

    Instructions for use ESTETICA E50 Life 4 Operation | 4.10 Using instruments Note For units with a BS selective holder, a service technician can set the vacuum stop function to either stop all suction hoses or only stop the spray mist suc- tion when the saliva ejector is simultaneously activated.
  • Page 100 Instructions for use ESTETICA E50 Life 4 Operation | 4.10 Using instruments ① Air button (A) ② Cannula ③ Gripping sleeve ④ Water button (W) ⑤ Ring blue Note The cannulas can be rotated 360 ▶ Remove the turbine from the holder.
  • Page 101: Using The Multifunctional Handpiece

    Instructions for use ESTETICA E50 Life 4 Operation | 4.10 Using instruments 4.10.4 Using the multifunctional handpiece CAUTION Risk of injury from touching the cheek with the handpiece. Irritation of the mucosa. ▶ Rotate the cannula of the handpiece into an operating position where there is no contact of the mucosa.
  • Page 102 Instructions for use ESTETICA E50 Life 4 Operation | 4.10 Using instruments Note Cannulas can be rotated by 360 The "on" time for the handpiece with heating is 5 minutes with a resting time of 3 minutes. Note If only the cold light is preselected (heater: off), the multifunctional hand- piece shines when it is removed from the holder.
  • Page 103 Instructions for use ESTETICA E50 Life 4 Operation | 4.10 Using instruments ▶ Press the air button ① and/or the water button ②. ▶ Press the "Handpieces" foot pedal. ð The light turns on. Replacing the lamp CAUTION Danger of injury from a hot valve body.
  • Page 104: Using The Piezoled

    To ensure proper function, the poles need to be in- serted correctly. ▶ Push the holder ③ forward and pull the defective KaVo MULTI LED lamp ② out of the socket. ▶ Insert new Kavo MULTI LED lamp (Mat. no. 1.007.5372).
  • Page 105: Using The Kl 703 Led / Kl 702 In Endo Mode (Optional Accessory)

    Incorrect transmission ratio. Damage from incorrect speed / incorrect torque. ▶ Only use KaVo 1:1 reducing shanks 20LH or 20LP or MASTERmatic LUX M20 L with 1:1 INTRA LUX head L68 B (Mat. no. 1.008.1834) or 3:1 IN- TRA head L66 B (Mat. no. 1.008.1831).
  • Page 106: Open Endo Mode

    Instructions for use ESTETICA E50 Life 4 Operation | 4.11 Using the KL 703 LED / KL 702 in ENDO mode (optional accessory) CAUTION Excessive torque. Injury or damage to instruments. ▶ Use root canal instruments in ENDO mode only.
  • Page 107: Change Settings In The Option Menu

    Instructions for use ESTETICA E50 Life 4 Operation | 4.11 Using the KL 703 LED / KL 702 in ENDO mode (optional accessory) Parameters in the display ① Direction of motor rotation ② Torque mode ③ Transmission factor ④ ENDO Mode ⑤...
  • Page 108 Instructions for use ESTETICA E50 Life 4 Operation | 4.11 Using the KL 703 LED / KL 702 in ENDO mode (optional accessory) The following parameters can be changed: ▪ Speed ▪ Torque ▪ Torque mode Select the parameter memory ▶...
  • Page 109 Instructions for use ESTETICA E50 Life 4 Operation | 4.11 Using the KL 703 LED / KL 702 in ENDO mode (optional accessory) Save the parameters in programs 1 to 6 using the "Program" key (press for 2 seconds). You can save after setting one parameter, or after setting all para- meters.
  • Page 110 Instructions for use ESTETICA E50 Life 4 Operation | 4.11 Using the KL 703 LED / KL 702 in ENDO mode (optional accessory) ð The motor starts by rotating clockwise (if not selected otherwise). The torque is limited to the set threshold. The speed reduces until it stops de- pending on the load.
  • Page 111: Leaving Endo Mode

    Instructions for use ESTETICA E50 Life 4 Operation | 4.12 Use pump for physiological saline solution (optional accessory) ▶ To stop this, release the foot pedal. ▶ Press the foot pedal ð The motor rotates to the right. 4.11.5 Leaving ENDO Mode ▶...
  • Page 112: General Information

    Instructions for use ESTETICA E50 Life 4 Operation | 4.12 Use pump for physiological saline solution (optional accessory) 4.12.1 General information Overview of saline pump ① Air nozzle holder ② Salt bag ③ Hose set sterile ④ Knurled screw ⑤ Pump ⑥...
  • Page 113 Instructions for use ESTETICA E50 Life 4 Operation | 4.12 Use pump for physiological saline solution (optional accessory) Note The hose set sterile (10 pcs.) (Mat. no. 1.012.4601) is designed for single use and must be changed after each application.
  • Page 114: Activating The Pump For The Respective Holder (Enabling) And Regulating The Pump

    Instructions for use ESTETICA E50 Life 4 Operation | 4.12 Use pump for physiological saline solution (optional accessory) 4.12.3 Activating the pump for the respective holder (enabling) and regulating the pump Requirement Treatment centre is turned on. The handpiece is connected to the pump via the pressure line.
  • Page 115: Changing The Coolant Container

    Instructions for use ESTETICA E50 Life 4 Operation | 4.12 Use pump for physiological saline solution (optional accessory) ▶ Gradually adjust the amount of saline solution using the metering head. Rotation in "+" direction: increases the quantity Rotation in "-" direction: reduces the quantity ▶...
  • Page 116: Attaching And Etaching The Pump

    Instructions for use ESTETICA E50 Life 4 Operation | 4.12 Use pump for physiological saline solution (optional accessory) 4.12.5 Attaching and etaching the pump Mounting the pump Note Make sure that the pump is mounted insulated with the plastic plate on the table housing or holder.
  • Page 117: Replacing The Hose Set Sterile

    Instructions for use ESTETICA E50 Life 4 Operation | 4.12 Use pump for physiological saline solution (optional accessory) ▶ Turn off the treatment centre. ▶ Removing the hose set sterile ②. ▶ Unplug electrical line ①. ▶ Unscrew the knurled screw under the pump.
  • Page 118 Instructions for use ESTETICA E50 Life 4 Operation | 4.12 Use pump for physiological saline solution (optional accessory) ▶ Remove the old hose set sterile and dispose of it. ▶ Make sure that the flexing system is horizontal (turn manually if necessary).
  • Page 119: Using The Comfortdrive 200 Xd/Comfortbase (Optional Accessory)

    200,000 rpm. It can be attached only to the KaVo COMFORTbase coupling. The hose of the KaVo COMFORTbase is part of the coupling and cannot be re- moved! Operation and changing the settings via the control element is identical to the INTRA LUX motor KL 703.
  • Page 120: Replacing The High-Pressure Bulb Of The Comfortbase

    Instructions for use ESTETICA E50 Life 4 Operation | 4.13 Using the COMFORTdrive 200 XD/COMFORTbase (optional accessory) 4.13.3 Replacing the high-pressure bulb of the COMFORTbase CAUTION Danger of burns from hot high-pressure lamp. ▶ Switch main device switch off. ▶ Let the COMFORTbase cool down after long use.
  • Page 121: Using The Usb Interface

    The treatment unit may be fitted with up to three USB ports. Camera interfaces are situated on the underside of the dentist element (T-table) or in the dentist element (S-table). Only the cameras approved/enclosed in the delivery by KaVo may be connected to these interfaces.
  • Page 122: Preparation Methods Din En Iso 17664

    Instructions for use ESTETICA E50 Life 5 Preparation methods DIN EN ISO 17664 5 Preparation methods DIN EN ISO 17664 Note The preparation methods can be found in the care instructions. 122 / 162...
  • Page 123: Accessories And Kits

    Instructions for use ESTETICA E50 Life 6 Accessories and kits | 6.1 Device 6 Accessories and kits 6.1 Device Description Description Water block DVGW with With a DVGW permit and electronic monitoring of integrated water germ the filling level of the cleaning container.
  • Page 124: Dentist Unit

    KL 701 KaVo COMFORTdrive 200 Dental handpiece for the high speed range up to 200,000 rpm. It can be attached only to the KaVo COMFORTbase coupling. Triple function handpiece Multifunctional handpiece featuring air, water, no heating, and no cold light.
  • Page 125: Safety Check - Test Instructions

    Note If several medical electrical devices (ME device) or electrical devices from several manufacturers combined into a system are connected to the KaVo dental unit, the manufacturer data contained in the instructions for use for all products subject to the safety checks must also be noted.
  • Page 126: Notes For Medical Electrical Systems

    7 Safety check - Test instructions | 7.1 Introduction Note KaVo offers a medical device book for keeping an inventory and recording es- sential master data on the medical device. The medical device book is only available in German (Mat. no. 0.789.0480).
  • Page 127: Components Of The Safety Check

    Instructions for use ESTETICA E50 Life 7 Safety check - Test instructions | 7.1 Introduction Note If it should be impossible to check single ME devices that are functionally connected to an ME system individually for technical reasons, the ME system must be checked as a whole.
  • Page 128: Notes On Repeat Testing

    Instructions for use ESTETICA E50 Life 7 Safety check - Test instructions | 7.2 Instructions for the safety check 7.1.6 Notes on repeat testing Note The value determined in these tests must be documented and evaluated to- gether with the measuring processes. The measured values may not over- shoot the specified values.
  • Page 129 Instructions for use ESTETICA E50 Life 7 Safety check - Test instructions | 7.2 Instructions for the safety check Without/with DEKAmat/CENTRAmat ▶ Take off the cover ② proceeding in upward direction. ▶ Release the rear cover ① below and remove it.
  • Page 130: Visual Inspection (Inspection By Examination)

    Instructions for use ESTETICA E50 Life 7 Safety check - Test instructions | 7.2 Instructions for the safety check Without/with DEKAmat/CENTRAmat 7.2.2 Visual inspection (inspection by examination) Check the following items in advance: ▪ Has the equipment of the ME device or the ME system been changed since the last inspection? ▪...
  • Page 131 Instructions for use ESTETICA E50 Life 7 Safety check - Test instructions | 7.2 Instructions for the safety check Visual inspection and appraisal of the medical device and accessories The following list is for exemplary purposes and makes no claim of being com- plete.
  • Page 132: Measurements

    Instructions for use ESTETICA E50 Life 7 Safety check - Test instructions | 7.2 Instructions for the safety check Attachment locations: nameplate and note, "Comply with the instructions for use" Control of the availability of the necessary documents ▶ Check if the required instructions for use and care instructions are available in the surgery.
  • Page 133 L & N on the power input board need not be disconnected. The adapter cable ② is included in the delivery of the KaVo measuring cable and is required for older treatment centres that are not equipped with an X2 connector.
  • Page 134 Instructions for use ESTETICA E50 Life 7 Safety check - Test instructions | 7.2 Instructions for the safety check ▶ Plug the second plug X2 of the KaVo measuring cable into the network card (X2). ▶ Insert the protective contact plug of the KaVo measuring cable into the safety tester.
  • Page 135 ACPs on the treatment centre No ACPs need to be connected to the protective conductor (PE) during the measurement on the treatment centre ESTETICA E50 Life, as all relevant parts are connected to the PE at the factory and are included in the test.
  • Page 136 Instructions for use ESTETICA E50 Life 7 Safety check - Test instructions | 7.2 Instructions for the safety check Note The integrity of the power supply cable, in particular the protective earth wire of the power cable must be ensured. As this is a fixed installation, the evalu- ation can be conducted by means of a visual inspection.
  • Page 137 Instructions for use ESTETICA E50 Life 7 Safety check - Test instructions | 7.2 Instructions for the safety check Scan the treatment centre with the test tip Measuring points on the device base ① Main switch holding plate ② Stand cover base plate ③...
  • Page 138 Instructions for use ESTETICA E50 Life 7 Safety check - Test instructions | 7.2 Instructions for the safety check Scanning the patient chair with the test tip ① Top part of the chair ② Base plate of chair base ③ Chair switched mode power supply COMPACTchair measuring points ①...
  • Page 139 Instructions for use ESTETICA E50 Life 7 Safety check - Test instructions | 7.2 Instructions for the safety check Scanning the operating elements with the test tip ① Dentist element S: Fixing screw on ② Dentist element TM: Fixing screw...
  • Page 140 Instructions for use ESTETICA E50 Life 7 Safety check - Test instructions | 7.2 Instructions for the safety check Operating light EDI/MAIA No measuring points need to be scanned on the operating lights EDI and MAIA. 140 / 162...
  • Page 141 Instructions for use ESTETICA E50 Life 7 Safety check - Test instructions | 7.2 Instructions for the safety check Touch monitor with test tip: ▶ Scan the measuring point ① with the test tip. ▶ Scan the measuring point ② after removing the display cover.
  • Page 142: Functional Test

    Instructions for use ESTETICA E50 Life 7 Safety check - Test instructions | 7.2 Instructions for the safety check Applied part leakage current - Alternative measuring method Limit < 5 mA (maximum value!) Protection class 1 WARNING Electrical power. Death or injury from electric shock.
  • Page 143 Instructions for use ESTETICA E50 Life 7 Safety check - Test instructions | 7.2 Instructions for the safety check The following list is for exemplary purposes and makes no claim of being com- plete. ▪ Functional test of the safety circuits (see diagram below) ▪...
  • Page 144 Instructions for use ESTETICA E50 Life 7 Safety check - Test instructions | 7.2 Instructions for the safety check Item Safety switch-off ac- LED on assistant LED on dentist ele- tuated element ment ③ Backrest Bracket on the foot ④...
  • Page 145 Instructions for use ESTETICA E50 Life 7 Safety check - Test instructions | 7.2 Instructions for the safety check Display LED Safety shut-down Bench support/seat cushion, backrest, lower chair programme Bendable part of seat (COMPACTchair only) Foot control Patient element...
  • Page 146: Assessment And Documentation

    ▶ Name, date and signature of test engineer There is a copy of a test report template at the end of the chapter on Safety Checks. KaVo recommends the use of this template. Note Following testing, repair or adjustment, it must be verified whether the ME equipment or ME system has been restored to the state that is required for the intended usage before it is employed once again.
  • Page 147 Instructions for use ESTETICA E50 Life 7 Safety check - Test instructions | 7.2 Instructions for the safety check Unit leakage current - replacement measurement Leakage current of applied part - replacement measurement 147 / 162...
  • Page 148: Appendix - Additional Measuring Sites

    Instructions for use ESTETICA E50 Life 8 Appendix - Additional measuring sites | 8.1 Additional scanning sites SL X in the protective conductor measure- ment 8 Appendix - Additional measuring sites Note Regarding accessories, which are not listed here, comply with the specifica- tions of the relevant instructions for use.
  • Page 149: Additional Measuring Sites Ap X For Eul/Epl Measurement

    Instructions for use ESTETICA E50 Life 8 Appendix - Additional measuring sites | 8.2 Additional measuring sites AP X for EUL/EPL measurement Ceiling adapter for operating light assembly kit ① Base plate for the ceiling adapter ② Surroundings of the protective con- ductor connector ③...
  • Page 150: Additional Connection Sites Acp X (Additional Earth Connections)

    Instructions for use ESTETICA E50 Life 8 Appendix - Additional measuring sites | 8.3 Additional connection sites ACP X (additional earth connections) Note The switch on the handpiece must be activated during the EPA measure- ment. Note Additional measuring points AP X need to be taken into consideration in the presence of accessories: e.g.
  • Page 151: Troubleshooting

    Instructions for use ESTETICA E50 Life 9 Troubleshooting 9 Troubleshooting Note In case of malfunctions, consult the separate instructions for the use and care of the individual instruments (such as the turbine, motor, camera, Sate- lec Mini LED, etc.). Malfunction...
  • Page 152 Instructions for use ESTETICA E50 Life 9 Troubleshooting Malfunction Cause Remedy ▶ Also refer to: Instructions for use for the The Satelec Mini LED Also refer to: Instruc- Satelec Mini LED does not work. tions for use for the Satelec Mini LED ▶...
  • Page 153 Instructions for use ESTETICA E50 Life 9 Troubleshooting Malfunction Cause Remedy ▶ Replace all O-rings of the MULTIflex coupling. Water in the return air O-rings of the MUL- filter. TIflex coupling are damaged. ▶ Exchange the amalgam container. A melody sounds.
  • Page 154 Instructions for use ESTETICA E50 Life 9 Troubleshooting Malfunction Cause Remedy ▶ Restart the CONEXIO PC. Camera images shows Electrical or electro- images only as black/ magnetic interference white images. by other equipment. ▶ Replace the camera in the holder and then...
  • Page 155 Instructions for use ESTETICA E50 Life 9 Troubleshooting Malfunction Cause Remedy ▶ Refill the Oxygenal container. Display shows: ID 67 The oxygenal container is empty. See also: 2 Servicing instructions ▶ Have a service performed. Display shows: ID 68 Call for service ▶...
  • Page 156: 10Data On Electromagnetic Compatibility According To En 60601-1-2

    The treatment unitESTETICA E50 Life is designed for operation in an environ- ment as specified below. The customer or user of the ESTETICA E50 Life should make sure that the device is used in an environment of the specified type.
  • Page 157: Recommended Safe Distance Between Portable And Mobile Hf Telecommunications Equipment And The Treatment Unit

    HF telecommunications equipment and the treatment unit The ESTETICA E50 Life is intended for use in an electromagnetic environment in which the HF interference parameters are controlled. The customer or user of the ESTETICA E50 Life can help prevent electromagnetic interference by main-...
  • Page 158: Immunity To Electromagnetic Interference

    The treatment unitESTETICA E50 Life is designed for operation in an environ- ment as specified below. The customer or user of the ESTETICA E50 Life should make sure that the device is used in an environment of the specified type.
  • Page 159 A site study should be considered to determine the electromagnetic environment in terms of stationary transmitters. If the meas- ured field strength at the site, at which the ESTETICA E50 Life is used, exceeds the compliance levels shown above, the ESTETICA E50 Life should be mon- itored to demonstrate proper function.
  • Page 160 Instructions for use ESTETICA E50 Life 10 Data on electromagnetic compatibility according to EN 60601-1-2 | 10.4 Immunity to electromagnetic interfer- In the frequency range of 150 kHz to 80 MHz, the field strength should be less than 3V V/m.

Table of Contents