Care instructions ESTETICA E30
2 Reprocessing methods according to DIN EN ISO 17664 | 2.5 Reprocessing of the assistant element
▶ Wipe-disinfect the instrument holder.

2.5.2 Reprocessing of the suction hose connector

▶ Pull-off the connectors ① and ② from the suction hoses.
▶ Unscrew the connectors ③ and ④ from the suction hoses.
Cleaning and disinfection of the suction hose connector
▶ Clean the disassembled parts under running water.
▶ Thermodisinfect the disassembled connectors ① and ② for at least 10
minutes at 93 °C (total cycle: approximately 20 minutes).
▶ Disinfect the connectors ③ and ④ by wipe disinfection.
Note
Connectors ③ and ④ cannot be subjected to thermodisinfection.
Sterilising the suction hose connector
Damage caused by sterilisation.
▶ Do not sterilise connectors ① and ②.
Note
Sterilise the suction cannulas according to the manufacturer's instructions.
▶ Sterilise connectors ③ and ④.
6
5
6
5
NOTICE
28 / 40
1
2
3
4