Overview Estetica E30 Reprocessing - KaVo ESTETICA E30 Care Instructions

Hide thumbs Also See for ESTETICA E30:
Table of Contents

Advertisement

2 Reprocessing methods according to DIN EN ISO 17664 | 2.2 General instructions for the reprocessing of KaVo
Manual drying
Automatic drying
Servicing
Control and functional test
Packaging
Sterilisation
Storage
Reprocessing items/medical
devices
Surfaces
Device
Cushion
Light
Lights, handles
Dentist element
Handle TM table / handle cup S table
Cannula triple-function handpiece
Holder pad/anti-slip mat
Handpiece holder
Instrument arm S table
Assistant element
Connection parts of the suction hose
Patient element
Tumbler holder
Tumbler filler
Porcelain spittoon bowl
▶ Supply dry, clean and uncontaminated compressed air according
to ISO 7494-2 only.
▶ Blow off the outside and inside with compressed air until water
drops are no longer visible.
Usually, automatic drying is part of the cleaning/disinfection cycle of
the washer disinfector and should not be carried out above 95 °C.
▶ Discard damaged products.
▶ Perform a visual inspection to check for damage and wear and
tear.
▶ Test products for function regularly.
▶ Use standardised packaging material. Sealing should not be un-
der tension.
▶ Use sterile goods packaging in order to ensure sterility.
All sterilisable products are temperature-resistant up to max. 138
°C.
KaVo recommends sterilisation in a steam steriliser (autoclave) ac-
cording to EN 13060 / ISO 17665-1, e.g. STERIclave B 2200 / 2200
P (KaVo) or Citomat K-Series (Getinge).
▶ Sterilise with a triple fractionated vacuum (B class steriliser);
hold for 4 minutes; 134°C ± 1 °C; overpressure: 2.13 bar.
▶ Reprocessed products should be stored protected from dust with
minimum exposure to germs in a dry, dark and cool place.
▶ Comply with the expiry date of the sterilised items.

2.2.1 Overview ESTETICA E30 reprocessing

Cleaning
manual
X
X
X
X
X
X
X
X
X
X
X
X
X
Disinfection
automatic manual
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
7 / 40
Care instructions ESTETICA E30
treatment centres
Sterilisa-
tion
automatic automatic
See also:
2 IfU KaVoLUX 540
LED / EDI / MAIA
X
X
X
X
X
X
X