Reprocessing Of The Instrument Arm S Table - KaVo ESTETICA E30 Care Instructions

Hide thumbs Also See for ESTETICA E30:
Table of Contents

Advertisement

2 Reprocessing methods according to DIN EN ISO 17664 | 2.4 Preparation of the dentist element
Sterilize the instrument holder
Damage caused by sterilisation.
▶ Do not sterilise the instrument holder.

2.4.3 Reprocessing of the instrument arm S table

Note
The swing-arm S-table (Mat. no. 10074755)) can be removed to facilitate
cleaning.
Cleaning and disinfection of the instrument arm S table
Damage caused by treatment in a washer disinfector.
▶ Do not sterilise the instrument arm S table in a washer disinfector.
▶ Take the handpieces off the hoses.
▶ Take the roller out of the instrument arm.
▶ Pull-off the instrument arm in an upward direction.
▶ Clean the instrument arm and roller under running water.
▶ Wipe the instrument arm S table with disinfectant.
Sterilisation of instrument arm S table
Damage caused by sterilisation.
▶ Do not sterilise the instrument arm S table.
NOTICE
NOTICE
NOTICE
11 / 40
Care instructions ESTETICA E30