Reprocessing Of The Spittoon Bowl - KaVo ESTETICA E30 Care Instructions

Hide thumbs Also See for ESTETICA E30:
Table of Contents

Advertisement

Care instructions ESTETICA E30
2 Reprocessing methods according to DIN EN ISO 17664 | 2.6 Reprocessing of the patient element
Cleaning and disinfection of the tumbler holder and
tumbler filler
▶ Take off the tumbler holder (Mat. no. 1.008.9291).
▶ Clean the tumbler holder under running water.
▶ Subject the tumbler filler to wipe disinfection.
Damage caused by treatment in a washer disinfector.
▶ Do not subject the tumbler holder and tumbler filler to treatment in a
washer disinfector.
Sterilisation of tumbler holder and tumbler filler
Damage caused by sterilisation.
▶ Do not sterilise the tumbler holder and filler.

2.6.3 Reprocessing of the spittoon bowl

Water in the device.
Material damage from leaking water upon actuation of the "Tumbler filler" or
"Bowl flush" keys.
▶ Make sure that the unit is turned off before removing the tumbler filler or
the spittoon bowl.
▶ Always check the correct seating of tumbler filler and spittoon bowl before
turning on the unit.
Cleaning and disinfection of the spittoon bowl
Damage caused by inappropriate cleaning agents.
▶ Do not use acid-containing, sand-containing, and form-forming cleaning
agents.
NOTICE
NOTICE
NOTICE
NOTICE
30 / 40