Filling The Intensive Germ Reduction Bottle - KaVo ESTETICA E30 Care Instructions

Hide thumbs Also See for ESTETICA E30:
Table of Contents

Advertisement

3 Replenishing and checking the disinfectant | 3.1 Replenishing KaVo OXYGENAL 6 when using the water bottle
▶ Squeeze the KaVo OXYGENAL 6 bottle until the germ reduction liquid rises
ð

3.1.1 Filling the intensive germ reduction bottle

Wrong germ reduction liquid or the dose is too high
Injury hazard
▶ Use KaVo OXYGENAL 6 only.
▶ Do not exceed the maximum concentration of 0.25% KaVo OXYGENAL 6.
▶ Proceed like for filling the water bottle, but use a different dose of KaVo
See also:
2 Filling the water bottle with the dosing facility, Page 36
▶ Fill the four-fold dose of KaVo OXYGENAL 6 (equivalent to 20 ml) in the in-
▶ Then fill-up the intensive germ reduction bottle with tap water.
above the red marker.
In response to over-filling, the excess quantity above the tick mark is aspir-
ated automatically to ensure exact dosing.
OXYGENAL 6.
tensive germ reduction bottle. Make sure that the spout of the KaVo OXY-
GENAL 6 dosing facility (nose) is situated right above the opening of the in-
tensive germ reduction bottle.
CAUTION
37 / 40
Care instructions ESTETICA E30