2 Reprocessing methods according to DIN EN ISO 17664 | 2.4 Preparation of the dentist element

2.4.7 Servicing the turbine return air filter

Cleaning the turbine return air filter
Note
The liquid collector must be checked weekly.
▶ If there is any liquid in the filter, drain the liquid and replace the O-rings of
the multiflex couplings.
▶ If necessary, replace the turbine return air filter (Mat. no. 02003098) and
clean the filter holder.
Note
The turbine return air filter is in the same position in the dentist element vari-
ant with S table.

2.4.8 Reprocessing KaVo ERGOcam One

Note
The camera handle cannot be sterilised or disinfected with heat.