KOTS Medium-Frequency Muscle Training
according to Kots
Fixed parameters
R = release time: 10 ms
Sine
Application: Muscle training

2.4.3 Current Modes of the Diagnostics Menu

I/T Curve Diagnosis
T
R
T
R
PHYSIOMED-Jubilee
Adjustable
T = Contraction 1 – 60 s;
manual triggering (manual release key)
R = release time between contractions 0 – 60
s
T = pulse time:
2 ms, 4 ms, 6 ms, 8 ms, 10 ms; auto
Pulse shape: sine, triangle, rectangle
Ramp (surge time): 4 modes
Modulation frequencies: 0 – 95 Hz
F
= Basic frequency: 2.0 - 9.5 kHz
bas
Triangle
T
T
Rectangle
Pulse shape: square, triangular
Pulse time T: 0.1 – 1000 ms
Release time R: 2 s, 3 s, 4 s, 5 s, manual