11 Accessories

Warning For safety reasons,
cessories. The use of other manufacturers' accessories is at the user's risk.
Warning Plate electrodes are subject to wear and tear. The degree of wearout depends on
the current mode, the frequency of use, and the intensity. The greater the galvanic
component of the application is, the quicker is the wearout of the electrodes.
Since poorly conducting electrodes can lead to skin impairment, we recommend
checking the electrodes regularly (at least once per quarter) using the Electrode test
pen (ref. no. 00412). Alternatively you can send the electrodes to our service de-
partment for an inspection, which is liable to pay a service fee.

11.1 Standard Accessories

Ref.-No.
Description
00179
Accessory tray
00451
Elastic velcro strap 6 x 80 cm
00453
Elastic velcro strap 10 x 125 cm
00776
Mains cable
00331
Manual release key
00185
Patient lead (4 connectors)
00476
Plate electrode EF 50 (set of 2)
00435
Viscose cover EF 50
00946
Operating Instructions (English)
00625
Short Introduction to Electrotherapy (English)
Warning An effective current density of 2 mA/cm² must not be exceeded! For this reason you
should always use electrodes of a sufficient size and be especially attentive when
attaching them.
For more detailed information on size and application of the electrodes in therapy refer to
our therapy booklet supplied.
PHYSIOMED-Jubilee
has to be used exclusively with original ac-
PHYSIOMED-Jubilee
Accessories