Treatment Procedure using Automatic Surge Current
(1) Attach the electrodes to the respective stimulation points. The position of the
patient must enable a free movement of the stimulation area.
(2) Depending on the indication, select the current modes
(3) Set the tension time T and the release time R. The release time R should be long
enough to give the nerve-muscle-unit enough recreation time.
(4) Select the ramp of the current rise.
(5) Select the modulation frequency.
(6) If necessary, alter the basic frequency.
(7) Activate the acoustic signal, if desired.
(8) Set the therapy time.
(9) If necessary, plug in the manual release key as therapy pause button.
(10) Slowly turn up the intensity with intensity control <6>). The therapy time
elapses.
Warning Take care that the electrodes are by no means touched after the cur-
rent is turned up!
(11) Signs of fatigue should stop the treatment (they may already occur after a few
muscle contractions)!
(12) After treatment, an acoustic signal will be issued.
Turn the intensity down to 0.
(13) Remove the electrodes.
Treatment Procedure using Manual Release Key
Surge current treatment by manual release key is advisable for individual therapy
methods in the field of sports, for intentional exercises according to Förster, as well
as for practising both co-ordination and nerve tracts. This is where the manual trig-
ger of the Surge current proves its value.
(1) Plug the manual release key into the manual release key socket <11> at the
rear face of the instrument.
(2) Attach the electrodes to the respective stimulation points. The position of the
patient must enable a free movement of the stimulation area.
(3) Depending on the indication, select one of the current modes
(4) Set the release time R to 1s and the tension time T to the left to the manual re-
lease key position.
(5) Select the ramp of the current rise.
(6) Select the frequency of modulation.
(7) If necessary, alter the basic frequency.
(8) Activate the acoustic signal, if desired.
(9) Set the therapy time.
(10) Slowly turn up the intensity with manual release key pressed (intensity con-
trol <6>) until the muscle contraction is sufficient. The therapy time elapses.