Stimulation Current Therapy
Warning Take care that the electrodes are by no means touched after the cur-
rent is turned up!
Warning When using the manual release key, release a pulse each time you turn
up the intensity!
(11) Release the button.
If the manual release key is pressed again, the current at FaS and HVS will in-
crease automatically and will keep the set intensity as long as the manual release
key is pressed. After releasing the button, the intensity will automatically be re-
duced in intervals.
(12) Signs of fatigue should stop the treatment (they may already occur after a few
muscle contractions)!
(13) After treatment, an acoustic signal will be issued. Turn the intensity down to 0.
(14) Remove the electrodes.
Note
If the intensity (intensity control <6>) is turned up during the
time of release, the following message will appear in the lower
status bar <4>:
Slowly turn down the intensity then. The message will disappear as
soon as the previously set value is reached again.
34
PHYSIOMED-Jubilee