Notes on Operation

3 Notes on Operation
3.1

Connection and Start-up

Connecting the unit
(1) Check whether the operating voltage of the device (see red window in the
mains module) <1> ) and the line voltage correspond.
(2) Plug the supplied mains lead firmly into the rear panel of the device (mains
module <1>) and connect it to the socket. A safe connection is established via a
grounding socket outlet with protective wire.

3.2 Device Start-up

(1) Make sure that the instrument is switched off.
(2) Turn intensity control <6> to 0.
(3) Insert the patient lead into the patient lead connector <10>.
(4) Switch on the device with the mains switch (mains module <1>).
The instrument will carry out an automatic self-test routine during which all
functions and output values are checked.
An acoustic signal is issued.
The instrument is ready now. You are in the start menu.
Warning Take care that the electrodes are by no means touched after the cur-
rent is turned up!
3.3

Self-test routine

(1) Turn intensity control <6> to 0.
(2) Switch the device off and on with the mains switch (mains module <1>).
The instrument will carry out an automatic self-test routine during which all
functions and output values are checked.
An acoustic signal is issued.
(3) The instrument is ready now. You are in the start menu.
22