Troubleshooting - PerkinElmer Lambda 365 Installation Instructions Manual

5-pos variable pathlength cell holder
Hide thumbs Also See for Lambda 365:

Advertisement

7. When calibration is finished, the following box appears. Select OK.
8. Select Save Result

Troubleshooting

W hen the 5-Pos Variable Pathlength Cell Holder does not m ove
1. Check the Connector status.
 Check if the 5-Pos Variable Pathlength Cell Holder connector is connected firmly to the Lambda
365.
W hen the intensity value is low
1. Recalibrate the 5-Pos Variable Pathlength Cell Holder.
 Recalibrate if the light beam does not go through the center of the cell holder hole.
2. Replace the lamp.
 If the intensity value is still low after the recalibration, the lamps of the light source for
measurement may be deteriorated and need replacing. Contact PerkinElmer.
to save the result. If the following message box appears, select OK.
11
09931462A