PerkinElmer Lambda 365 Installation Instructions Manual page 8

5-pos variable pathlength cell holder
Hide thumbs Also See for Lambda 365:

Advertisement

8. ① Click the Accessory tab and ② Select Variable Pathlength Multi-Cell on Accessory
Type, and then click the ③Method Setup button to open the Multicell window.
9. The Multicell window will be shown. Using the
available cells list. Check (√) Multi Baseline, if it is required to use.
Available Cells: Indicate cells that are available for measurement.
a.
b.
Using Cells: Shows the cell positions which are selected for measurement. Remove cells
using
Multicell Options
c.
Multi Zero: If checked (√), zero will be measured at all the selected cell positions. If not,
zero will be only measured at the 1
Multi Baseline: If checked (√), the baseline will be measured at each selected cell. If not
checked, the baseline is only measured at the 1
NOTE:
To prevent inadequate baseline (zero) measurement because of the difference in transmittance among
cuvettes or any other reasons, Multi Baseline (Multi Zero) measurement is recommended.
10. Click OK.
11. After setting parameters for Experiment, Lamp, Internal Reference and Scan Setup, click
Apply and then select OK.
12. After placing the reference solution into both the selected Using Cell positions and the reference
cell position of the Multicell holder, select the Baseline or Zero buttons to perform the Baseline
function.
NOTE:
The baseline is required to run only one time when the first measurement is done, not every
measurement. However, it should be done whenever the system is power off/on, the wavelength is
changed, SBW is changed or Reference sample is changed.
key and switch positions using
st
cell position among the using cells.
8
button, select cells to be used among the
and
keys.
st
cell position among the using cells.
09931462A