PerkinElmer Lambda 365 Installation Instructions Manual page 10

5-pos variable pathlength cell holder
Hide thumbs Also See for Lambda 365:

Advertisement

3. The functions of the Multicell calibration are shown as follows.
Command
MultiCell Position
Multicell Move
Multicell Calibration
Stop
Save Result
Reset Multicell
4. Click on Reset Multicell
5. Select MultiCell Calibration
samples from the 5-Pos Variable Pathlength Cell Holder (Empty the multicell holder).
6. Select OK, and then the Multicell calibration will start. The current process of calibration will be
shown in the main window.
Function
Show saved steps about each cell position of the Multi-Cell.
Used for moving Multi-Cell position as clicking buttons
.
Used for moving Multi-Cell position using
entered step.
Used to perform the Multi-Cell calibration.
Used to stop the Multi-Cell calibration.
Used to save the calibrated result.
Used to move to '0' step of Multi-Cell position.
to format the Multi-Cell steps. Click OK.
, and then the following dialog box will appear. Remove all
10
09931462A
buttons by