Lam bda 465 Fiber Optic Probe I nstallation I nstructions
This instruction sheet describes the installation of this accessory which is used with the Lambda 465
Spectrophotometer.
NOTE:
Read these instructions before you install this accessory.
Contacting P erk inElm er
Supplies, replacement parts, and accessories can be ordered directly from PerkinElmer, using the part
numbers.
See our website:
http://perkinelmer.com
PerkinElmer's catalog service offers a full selection of high-quality supplies.
To place an order for supplies and many replacement parts, request a free catalog, or ask for information:
If you are located within the U.S., call toll free 1-800-762-4002, 8 a.m. to 8 p.m. EST. Your order will be
shipped promptly, usually within 24 hours.
If you are located outside of the U.S., call your local PerkinElmer sales or service office.
I ntroduction
The probes provide a convenient way of performing absorption or transmission measurements by dipping
or inserting the probe end into the medium or environment. This accessory is especially useful for
embedding into process streams for real-time, on-line sample monitoring.
Features
Easy to install
Dip Probe for measuring liquid samples without using a cell
Figure 1 Lambda 465 Fiber Optic Probe [P/N: N4104012]
with Fiber Optic Probe [Part No: N4101048]
09931297D
PerkinElmer, 710 Bridgeport Avenue,
Shelton, CT 06484-4794, U.S.A
Produced in the USA.

Advertisement

loading

Summary of Contents for PerkinElmer Lambda 465

  • Page 1 09931297D Lam bda 465 Fiber Optic Probe I nstallation I nstructions This instruction sheet describes the installation of this accessory which is used with the Lambda 465 Spectrophotometer. NOTE: Read these instructions before you install this accessory. Contacting P erk inElm er Supplies, replacement parts, and accessories can be ordered directly from PerkinElmer, using the part numbers.
  • Page 2: Dimensions And Specifications

    Breakout Length (BOL) 18 ~ 23cm Probe Length (PL) 15.2 cm 150(W) X 100(H) X 84(L) Dimension (mm) Fiber Coupling Mount 0.75 Weight (Kg) Description Configuration of the Fiber Optic Probe Figure 2 Lambda 465 Fiber Optic Probe include Fiber Coupling mount...
  • Page 3: Installation

    09931297D I nstallation 1. Disassemble the existing cell holder. Figure 3 Location of the two knob bolts 2. Install the fiber optic probe by fastening the knob bolts. Figure 4 Installing the fiber Optic probe 3. Connect the communication cable and the power cord, and then turn on the power of the Lambda 465.
  • Page 4 09931297D 4. Place the probe into the liquid sample. Figure 5 Placing the probe in the liquid M easurem ent NOTE: Start the sample measurement after warming up at least 20 minutes. Launch the UV Lab software. Choose the Fiber Module in the Accessory Type. After dipping the probe into the blank solution, select the intensity mode, and then click Blank.
  • Page 5: Troubleshooting

    09931297D Select the Absorbance mode and measure blank and sample. Before measurement, make sure that bubble does not exist at the mirror side of the probe. Clean the probe with D.I water after the measurement. Troubleshooting W hen the I ntensity value is low Make sure that bubble does not exist at the mirror side of the probe.
  • Page 6 09931297D...