M easurem ent
Before executing System Self Test, the 5-Pos Variable Pathlength Cell Holder has to be installed first.
NOTE:
Otherwise, the instrument can be damaged electrically and do not operate properly.
1. Install the 5-Pos Variable Pathlength Cell Holder referring to chapter Installation.
2. Close the sample compartment cover and turn on the instrument.
CAUTION
ATTENTION
NOTE:
Start the System Self Test after warming up the system for at least 20 minutes.
3. Double-click on the UV Express software folder and select one of the modes for starting.
4. The following message box will be displayed. Empty the cell holder and close the lid firmly. Click
OK.
5. System Self Test starts. Click OK after finishing the System Self Test.
6. Click New to open a new window. Select Experiment Type and click OK.
7. Open Method in the main menu.
Do not start the UV Express software before finishing the initialization of grating.
Ne dém arrez pas le logiciel UV Express avant de terminer l'initialisation du
réseau.
7
09931462A