M easurem ent
Start the System Self Test after warming up the system for at least 20 minutes.
NOTE:
1. Double-click on the UV Express Software folder and select one of the mode for starting.
2. The following message box will be shown. Empty the cell holder and close the lid firmly. Click OK.
3. System Self Test stars. Click OK after finishing the System Self Test.
4. Select New to open a new window. Select Experiment Type and select OK.
5. Open Method in the main menu.
6. Perform multi-cell calibration according to the 'Calibration of Multi-Cell Position' chapter. (p 15)
Calibrate the beam position of the Changer 8-Cell for Dissolution whenever it is installed or beam position
NOTE:
is incorrect.
7. After completing the multi-cell calibration, remove the Plate for Calibration.
Figure 19. Removing the Plate for Calibration
9
09931465A