19. After the measurement is finished, the results are displayed in the result window. Save or print
the results as required.
System Self Test m ethod w hen using the Lam bda 365 w ith the
Changer 8-cell for Dissolution
NOTE:
When the Lambda 365 system is powered on again after turned off, the System Self Test need to be
performed with the cell holder empty. If you want to have the System Self Test done with a Changer 8-
cell Water Jack for Dissolution and manifold installed, follow the procedures below.
NOTE:
If you re-install the changer 8-cell water jack for dissolution after removing it, you need to perform the
multi-cell calibration referring to section 'Calibration of Multi-Cell Position' (page 15).
1. Insert cells for each position from 2 to 8.
Figure 25. Inserting cells from 2 to 8
2. Empty the No.1 and reference cell position, and place those two cells into the flow cell storage
block inside the cell compartment.
Figure 26. Inserting No.1 and reference cells into the flow cell storage block
3. Turn on the instrument.
13
09931465A