PerkinElmer Lambda 365 Installation Instructions Manual page 15

Changer 8-cell for dissolution
Hide thumbs Also See for Lambda 365:

Advertisement

Figure 28. Inserting No.1 and reference cells into the cell holder
8. And then click New to open a new window.
9. Open Method in the main menu.
10. Measure the sample according to the procedure 13-19 of Measurement section. (page 11)
Calibration of M ulti-Cell P osition
Calibrate the beam position of the Changer 8-Cell for Dissolution when the Multicell is installed for the first
time or beam position is incorrect.
1. Execute Scan Setup mode and ① click the Accessory tab of Method window. ② Select Multi-
cell on Accessory Type, and then click the ③Method Setup button to open the Multicell
window.