PerkinElmer Lambda 365 Installation Instructions Manual page 12

Changer 8-cell for dissolution
Hide thumbs Also See for Lambda 365:

Advertisement

14. Using the
Baseline or Multi Zero, if it is required to use. And then click OK.
a.
Available Cells: Indicate cells that are available for measurement.
Using Cells: Shows the cell positions which are selected for measurement. Remove cells
b.
using
Multicell Options
c.
Multi Zero: If checked (√), zero will be measured all the selected positions. If not, zero will
be only measured at the 1
Multi Baseline: If checked (√), the baseline will be measured at each selected cell. If not
checked, the baseline is only measured at the 1
To prevent inadequate baseline (zero) measurement because of the difference in transmittance among
NOTE:
cuvettes or any other reasons, Multi Baseline (Multi Zero) measurement is recommended.
15. After setting parameters for Experiment, Lamp, Internal Reference and Scan Setup, click
Apply and then select OK.
16. After placing the reference solution into the selected Using Cell positions and the reference cell
position of the Changer 8-Cell for Dissolution, click the Baseline or Zero button to perform the
Baseline or Zero function.
Baseline is required to run only one time when the first measurement is done, not every measurement.
NOTE:
However, it should be done whenever the system is powered off/on, the wavelength is changed, SBW is
changed or Reference sample is changed.
17. Place the reference into the reference cell position and sample into the selected Using cell
positions
18. Select the Sample button to measure.
button, select cells to be used among the available cells list. Check (√) Multi
key and switch positions using
st
cell position among the using cells.
12
and
keys.
st
cell position among the using cells.
09931465A