Ecg Synchronisation During The Treatment - IGEA Cliniporator EPS02 User Manual

Table of Contents

Advertisement

ECG Synchronisation during the Treatment

5.10
Only in case of ECT treatments, in the treatment screen (Figure 43) the icon
symbol present in the electrode selection window and an explanatory message below it inform
the operator about the synchronisation status with the ECG signal.
The symbols are described below:
• Green background and "ECG sync" writing: ECG enabled and working
• Yellow background and "Disabled" writing: ECG disabled
• Red background and "ECG Noisy" writing: ECG enabled but with disturbed signal
• Red background and "ECG Lost" writing: ECG enabled but signal not detected
• Red background and "ECG BAD" writing: ECG enabled but with incorrect signal
In all the different phases of the treatment, the ECG status is monitored and for every status
variation the corresponding symbol appears.
In case the icon
appears with a red background, the detection of the pressure of the ARM
pedal will be disabled and it will not be possible to deliver the treatment.
During application of the treatment, an anomaly on the ECG signal causes the temporary
suspension of the treatment itself. If an "ECG Noisy" signal is detected, the software
automatically resumes treatment delivery if the synchronization signal is restored within 3
seconds, but informs the user of the disturbance by displaying a message. If the ECG signal
remains stable for at least 5 seconds after the noise is detected, the message will disappear,
otherwise it will be displayed to indicate the presence of a disturbed ECG signal during which
the CLINIPORATOR will try, if possible, to continue the treatment delivery. In the event that the
disturbed signal protract over time and in other cases of abnormal ECG, the software proceeds
to open the dialog box of Figure 51 with the following indications:
message related to the detected anomaly ;
timer indicator with a 2 minute reverse count ;
"Resume" button
40

Figura 50 – Red ECG icon
and "Abort" button