Cliniporator model EPS02 – User manual
The pedal pulse (Pulse) green pedal in Figure 2 must be activated within 10 seconds after
pressing the pedal activation (Arm). Once the pedal has been pressed and the delivery of the
treatment starts, the system sends an acoustic signal and the information about the system
status shown in the status frame is updated (Figure 46 ).

Figure 46 - activation pedal pressed: ongoing treatment
In case the pedal pulse is not pressed within 10 seconds, the treatment procedure will
automatically be cancelled. If any treatment is delivered after 5 minutes the capacitors will
automatically be discharged. In this case use the charge button (Figure 43 ) or press the Arm
pedal to recharge them and restart the treatment.
At the end of the treatment, the system sends a second acoustic signal and, as shown in Figure
47, inside the frame
voltage and current waveforms are shown and the treatment counter is
increased
.



Figure 47 - Treatment screen: delivered treatment