Cliniporator model EPS02 – User manual
In the case of a pen drive that is not found or
faulty, the message in Figure 17 appears,
confirming that the shutdown procedure of
Figure 16 is activated.
Figura 17 – Messaggio di warning per pen drive non trovata
If an error is detected during the saving of the
data, the message of Figure 18 is shown
informing the user of the failure to save and
the possibility of removing the USB support.
Figura 18 – Messaggio di errore nel salvataggio
In the sole case in which the error diagnosed
is related to the automatic recognition device
of the electrode, the message shown in Figure
19 appears and the manual recognition
procedure is enabled.
Pressing the confirm button allows you to
operate exclusively in manual mode, as
described in paragraph 5.5.
The self-test resumes from the point where it
was interrupted.
Figura 19 – Errore sul circuito RFID
In section 7.2, a complete list of error messages is shown
.
WARNING
In case of automatic turn off during the autotest, wait for at least 10 seconds before
restarting the device by pressing the turn on button. This will avoid the system to detect
power supply errors in the selftest phase shown by the message "power supply failure".
22